P/N 9-500-0255
5-1
Service Manual
Chapter 5: Maintenance—General
Introduction
Chapter 5: Maintenance—General
1.0
Introduction
This chapter provides general information to aid the Service Engineer in maintaining the
Affinity. The information covered in this chapter is as follows:
•
Preventive Maintenance (Section 2.0)
•
Troubleshooting (Section 3.0)
2.0
Preventive Maintenance
The Affinity Mammography System recommended preventive maintenance procedures are
identified in Table 5-1. In addition, each time the Affinity Mammography System is down for
service:
1. The unit should be inspected for:
•
Cleanliness,
•
Loose or missing panels or covers, and
•
Loose or frayed wires or cables.
2. The Alert Codes Log should be checked for existing fault conditions as per Section 3.2.
3. Correct any problems found prior to returning the unit to operations.
Upon removing and replacing major components within the Affinity, any recommended
calibration or performance check procedures must be performed (as required). The
recommended frequency of performance codes used in Table 5-1 are defined as follows:
•
A = Annual
•
S = Semi-Annual
•
M = Monthly
•
W = Weekly
•
D = Daily
•
AR = As Required
Содержание AFFINITY Series
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ...Service Manual for Affinity Series Mammography Systems Part Number 9 500 0255 Revision 005 ...
Страница 6: ......
Страница 20: ...xiv P N9 500 0255 Service Manual Preface Acronym List ...
Страница 116: ...4 60 P N9 500 0255 Service Manual Chapter 4 Calibrations and Performance Checks Maintenance Procedures ...