  Instructions for use

  |  Elexo XXL Shower-toilet chair

Contact the technical service or Lopital’s service department . E-mail: info@lopital .nl 

4.4. On / off switch

On the right hand side of the cover cap, there is the on / 
off switch 

(See figure 20)

 . If the chair is out of use for an 

extended period, you are advised to turn it off . This will 
prevent any unauthorised use of the chair .
 

5. Cleaning and maintenance

The Elexo XXL must be cleaned after every use, with water and regular household 
cleaning agents . PUR integrated foam components with IMC coating should be rubbed 
clean and disinfected . Do not use any abrasive or other aggressive cleaning products . 
If necessary, the PUR components can be disinfected with a solution of 70% ethanol or 
propanol . 

The time the products must be left on the components has a significant impact upon 
the lifespan of the PUR components . As a result of the material properties of PUR, 
contact with sharp objects must be avoided . This could lead to damage or tears in 
the material . Finally, we would like to point out that non-critical parts of the PUR 
components must also be cleaned and disinfected .

Check the wheels regularly for hairs, dust and soap residues and remove when 
necessary . Maintenance and repairs to the Elexo XXL may only be carried out by 
authorised personnel .

Figure 20 On / off switch .

Содержание Elexo XXL

Страница 1: ...Instructions for use Elexo XXL Shower toilet chair ...

Страница 2: ...2 Instructions for use Elexo XXL Shower toilet chair Instructions for use Elexo XXL 5100 5300 Shower toilet chair Basic UDI DI 872025610316551005300JX ...

Страница 3: ...pital nl info lopital nl All of the aids produced by Lopital fulfil the following norms for medical products Class 1 MDR EU 2017 745 Explanation of symbols in instructions for use and on labels Note Follow the instructions for use Applied parts Read cleaning and disinfecting protocol Direct current 24V DC Antibacterial Class 2 EMC Temperature Date of production Air pressure Maximum load Humidity l...

Страница 4: ...rmrests 13 3 5 Moving upwards downwards 13 3 6 Toilet function 14 3 6 1 In combination with toilet 14 3 6 2 Use of a bedpan or bucket 14 3 7 Removable seat 15 3 8 Charging batterie 15 4 Safety functions 17 4 1 Emergency stop 17 4 2 Sleep setting 17 4 3 Emergency lowering 17 4 4 On off switch 18 5 Cleaning and maintenance 18 6 Environmental conditions 20 7 Removal of parts 20 8 Technical specificat...

Страница 5: ...in several tests Germs were demonstrably reduced in a short time to 99 99 The coating only affects bacteria that are found on the surface and not on airborne micro organisms The coating protects users or others from disease causing bacteria germs viruses or other harmful organisms On a dirty surface the coating loses its antibacterial properties So proper cleaning and disinfection is still a must ...

Страница 6: ...with more than 220 kg 485 lbs The Elexo XXL is only suitable for moving and caring for one person Only use original parts supplied by Lopital The use of components chargers cables actuators batteries etc that have not been supplied by Lopital may result in increased emissions or decreased immunity of the Elexo XXL Use the Elexo XXL in a clean and tidy working environment Only verifiably authorised...

Страница 7: ...XL shower toilet chair should only be operated by competent personnel who have experience in working with the user target group and have watched the instruction training videos on the Lopital website The Elexo XXL has been developed and produced for use in a professional healthcare environment Making changes to the contruction may affect the safety of the Elexo XXL Furthermore the liability and gu...

Страница 8: ...8 Instructions for use Elexo XXL Shower toilet chair Swivel wheel Hinge cover Frame Scissor mechanism Back support Charger Figure 1 Activating Armrest Seat Calf support Footrest frame Footrest ...

Страница 9: ...Elexo XXL Shower toilet chair Instructions for use 9 ...

Страница 10: ... contact with the user s while the Elexo XXL is in use 2 2 Activating The Elexo XXL must be activated before use which can be done by pressing any button on the manual control for 2 seconds See figure 1 It may occur that the Elexo XXL goes into sleep mode after a period of time To deactivate the sleep mode press any button on the hand control for 2 seconds Activate the Elexo XXL by pressing any bu...

Страница 11: ...raph 3 1 Wheel blocking Step 6 The client can take a seat in the chair Support the client if necessary Step 7 Place both feet on the footrests and calves against the calf supports See paragraph 3 3 Footrests Step 8 Close the armrests once the client has taken a seat in the chair See paragraph 3 4 Armrests Step 9 Notify the client about the processes you will be carrying out Step 10 Take the chair ...

Страница 12: ...t be pushed upwards again 3 3 Footrests Before the client takes a seat in the Elexo XXL the footrests must be raised slightly and turned outwards There is now sufficient room in front of the chair and the client can easily take a seat Once the client has taken a seat the carer or client raises one leg the carer then turns the footrest forwards until it falls into the lock The client can then rest ...

Страница 13: ...upport The armrest must not be used as a lever arm e g when moving the chair over an obstacle or as a push pull bar You must not sit on it either This could damage the Elexo XXL and or present a hazard 3 5 Moving upwards downwards Moving upwards If the client is seated in the chair first apply the brakes to the wheels and then move the chair upwards Move the chair upwards by pressing the button th...

Страница 14: ...ownwards to achieve the right position The Elexo XXL can now be moved backwards over the toilet Lower the Elexo XXL if necessary to achieve the right position The Elexo XXL does not fit over every toilet 3 6 2 Use of a bedpan or bucket If the Elexo XXL is being used as a toilet chair the special guides under the seat can be used to accommodate a stainless steel plastic bedpan or bucket The bedpan ...

Страница 15: ...nd the frame 3 8 Charging battery The Elexo XXL is an electrically driven high low shower toilet chair and works on a battery The charging point is located on the front of the chair on the blue cover on the underside of the device See figure 15 If the battery is almost completely empty of charge the orange light on the manual control will come on this means that the battery needs to be charged If ...

Страница 16: ...will not cause any damage Disconnect the Elexo XXL from the charger Once activated the Elexo XXL is now ready for use See paragraph 2 2 Activating LED not on Chair is in sleep setting and cannot be used Green LED is on Chair is ready for use Orange LED is on Battery needs charging or battery is partially charged Orange LED blinks Emergency stop has been activated Red LED blinks Battery must be cha...

Страница 17: ...ep setting After around an hour the chair will fall into a sleep setting and the battery indication LEDs on the manual control will go out In order to use the chair again it must be activated See paragraph 2 2 Activating 4 3 Emergency lowering If the controls stop working and the chair is in a raised position it can be lowered using the emergency lowering button Check first to make sure the chair ...

Страница 18: ...bed clean and disinfected Do not use any abrasive or other aggressive cleaning products If necessary the PUR components can be disinfected with a solution of 70 ethanol or propanol The time the products must be left on the components has a significant impact upon the lifespan of the PUR components As a result of the material properties of PUR contact with sharp objects must be avoided This could l...

Страница 19: ...tery at least once a week Annually According to the regulation MDR EU 2017 745 medical devices a chair must be maintained according to the manufacturer s guidelines This process will preferably be carried out by Lopital or another authorised body Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance For more information ab...

Страница 20: ...tions during use 7 Removal of parts All parts that are replaced or removed can be sent back to Lopital BV You may also offer the parts to the engineer for destruction We will ensure that all parts are processed in an environmentally friendly manner If you take care of disposal of the product yourself you must take account of the fact that batteries and chargers must be disposed of as small chemica...

Страница 21: ...d steel tubes and plates that are equipped with a two layer powder coating Hinge points Stainless steel cover caps made of plastic V0 Wheels Plastic Ø 12 5 cm 4 9 inch double blocking facility Body support PUR polyurethane Weight 71 kg 165 lbs Maximum permitted load 220 kg 485 lbs 8 2 Drive Drive 2 x actuator 24V DC Push Only 8000N Battery 2 x battery capacity 7Ah Charger 110 240AC 50 60Hz max 0 3...

Страница 22: ...22 Instructions for use Elexo XXL Shower toilet chair ...

Страница 23: ......

Страница 24: ... Electromagnetic Environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be made of wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply lines 1...

Страница 25: ... distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnet...

Страница 26: ...er of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 4 P 0 01 0 12 0 12 0 24 0 1 0 38 0 38 0 76 1 1 2 1 2 2 4 10 3 8 3 8 7 6 100 12 12 24 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicab...

Страница 27: ...e information visit our website www lopital com Lopital Nederland B V Laarakkerweg 9 5061 JR Oisterwijk Postbus 56 5060 AB Oisterwijk Tel 31 0 13 5239300 E mail info lopital nl www lopital nl User5300 02 2022 ...

Отзывы: