6.  Close doors and relatch by pressing down on the 

white latches so that the assembly looks like that 
shown in figure 5. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
7.  Place stirring bar, if desired, into bottom of reservoir 

in stirring corral. 

 

3.2 

Running the Gel 

1.  Place core assembly into lower reservoir. The anode 

(red) and cathode (black) electrodes are color-coded 
on both the core/cassette assembly and lower 
reservoir. Ensure the red dot on the cassette 
assembly is on the same side as the red receptacle 
on the lower reservoir. Fill core upper reservoir with 
freshly prepared buffer (~ 190 ml).

 

NOTE:  If any buffer is spilled into banana jack 
receptacle in lower reservoir, dry completely 
using compressed air! Failure to do this will 
result in accelerated banana jack corrosion. 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Banana jack 

receptacles

Figure 6 

Figure 4 

 
2.  See section 5.3 for recommended buffers.  

Pour only enough freshly prepared buffer into 
lower chamber so that the final buffer level is 
just below bottom of sample wells.  Using a 
pipette or syringe, thoroughly flush out the 
wells.   

3.  Load samples.  If outer lanes do not contain 

sample, it is recommended that you run 
standards and/or fill outer lanes with loading 
buffer to reduce smiling and wrap-around 
effects. 

 
4.  Attach safety cover and turn on magnetic stirrer      

(if desired).   

 
 
 
 

Figure 5 

 
 
 
 
 
 
 
 
 
 
 

Figure 7 

 
 
 
5.  Connect the leads to the power supply, matching the 

color-coded red-to-red and black-to-black. 

See 

Section 5.1 for recommended power conditions. 

Begin separation by electrophoresis.  Figure 7 
shows the closed unit ready for electrophoresis. 

 
 

 

Содержание PAGEr Minigel Chamber

Страница 1: ...ct proves defective during this period Lonza Rockland Inc will repair or replace it at our option free of charge if returned to us postage prepaid This warranty does not cover damage in transit damage caused by carelessness misuse or neglect normal wear through frequent use damage caused by solvent corrosion damage caused by improper handling or user alteration nor unsatisfactory performance as a ...

Страница 2: ...running single gels not shown 2 adaptors for running 9 cm gels not shown Section 3 Instructions for Electrophoresis 3 1 Preparing the Electrophoresis Unit 1 Place unit in authorized work area Remove safety cover from the assembled unit by simultaneously pressing down on white push pins with your thumbs while lifting up on yellow safety cover with your fingers Do not remove safety cover by pulling ...

Страница 3: ...mpressed air Failure to do this will result in accelerated banana jack corrosion Banana jack receptacles Figure 6 Figure 4 2 See section 5 3 for recommended buffers Pour only enough freshly prepared buffer into lower chamber so that the final buffer level is just below bottom of sample wells Using a pipette or syringe thoroughly flush out the wells 3 Load samples If outer lanes do not contain samp...

Страница 4: ...X Towbin 1X tris glycine buffer 20 Methanol buffer for 10 minutes b Cut a piece of PVDF or nitrocellulose membrane to fit the gel Do Not Touch Membrane with Bare Hands Wear gloves c Soak membrane in 50 methanol for 30 seconds d Soak membrane in water for 5 minutes e Soak membrane 5 minutes in 1X Towbin buffer until it no longer floats Do Not Allow Membrane to Dry f Cut 1 piece of filter paper 1 mm...

Страница 5: ...nditions will vary according to the number and type of gels used buffer conditions employed power input and the general goal of the experiment Refer to section 5 4 for in depth discussions on practical and theoretical approaches to protein gel electrophoresis Using standard SDS PAGE buffer systems see section 5 3 For two 1 0mm thick gels at room temperature use the following conditions at constant...

Страница 6: ... brief rinse with DEPC water is sufficient after a thorough wash followed by a quick rinse in 70 ethanol 6 2 Maintenance The following inspection and maintenance procedures will help maintain the safety and reliable performance of the PAGEr Minigel Chamber Replacement parts can be requested by calling 1 800 638 8174 Banana plugs and power cords should be inspected regularly If the banana plugs bec...

Отзывы: