14
Replacing the Fuse
For additional fuses, call Lonza Scientific Support or e-mail
To replace the fuse:
1. Turn off the main power switch at the rear of Power Supply and detach the power
cord.
2. Open the fuse compartment located inside the Power Entry Module by inserting a
small flat blade screwdriver into the slot below the ON/OFF switch. Turn the
screwdriver to gently pry open the fuse compartment.
NOTE
: The fuse compartment will not open with the power cord in place.
3. Pull the fuse holder out of the compartment and inspect the fuse. If the fuse is burned
or there is a break in the fuse element, replace the fuse with the identical type of fuse
(
T2AH250V
) as provided in the fuse holder (see figure below).
4. Place the fuse holder back into the compartment.
5. Snap the cover closed.