LRF-3000P Ultrasonic flowmeter
- 39 -
Temperature Sensitivity Settings
When the temperature difference is less than the
sensitivity you set, the energy would not accumulate
and the adjustable range is 0~20
o
C. Factory setting is
0.2
o
C by default.
M24
Temperature
*R
Source
Sensitivity
SHC
Value
0.20
℃
Specific Heat Capacity Settings:
0. CJ128SHC (National Standard)
1. Others
M24
Temperature
*R
Source
Sensitivity
SHC
Options
0.CJ128
Others
4.2
KJ/m³
℃
M25
Power Cut Compensation
Power cut compensation is a function to calculate the
current value during the outage based on the average
flow before power cut, so as to compensate the total
flow.
You could choose to enable/disable the function.
M25
PowerDown COMP
*R
Options
0.ON
M26
K Factor
Each flowmeter would be calibrated before leaving
factory to get its meter coefficient, which we call the K
coefficient.
You also could calibrate the flowmeter in service on
site through M26 menu.
M25
K Fctor
*R
Value
1.00