If the actual volume is known in advance, enter the calibration inter face and
input the actual volume directly.

Disp Vol

2.02mL

Disp T.

60.5"

Disp T.

60.5

Act Vol

2.020 mL

"

Act Vol

Flow 2.043mL/min

:

Pause T.

"

1.0

Pause Time

"

1.1

Flowrate

Dispensing

Test T.

0.5 ’

Cal Vol

132.3 L

μ

Test T.
Cal Vol

132.3 L

μ

30

"

2.02 mL

1.System Setting

2.Control Mode

3.Disp Setting

4.Calibration

1.System Setting

2.Control Mode

3.Disp Setting

4.Calibration

1.System Setting

2.Control Mode

3.Disp Setting

4.Calibration

Flow 2.043mL/min

:

Pause T.

"

1.0

Flow 2.043mL/min

:

Pause T.

"

1.0

Test T.

0.5 ’

Cal Vol

132.3 L

μ

Test T.

0.5 ’

Cal Vol

132.3 L

μ

Test T.
Cal Vol

133.0 L

μ

30

"

1.System Setting

2.Control Mode

3.Disp Setting

4.Calibration

Flowrate

Dispensing

Test T.

0.5 ’

Cal Vol

132.3 L

μ

Test T.

0.5 ’

Cal Vol

132.3 L

μ

Act Vol

133.0 L

μ

Flowrate

Dispensing

Disp T.

60.5

Act Vol

2.040 mL

"

1.System Setting

2.Control Mode

3.Disp Setting

4.Calibration

Flowrate

Dispensing

Disp Vol

2.02mL

Disp T.

60.5"

Flow Rate: Adjusting the dispensing flow rate can change the dispensing time.

Press and turn

to highlight dispensing flow rate line. Press

to enter

next inter face, turn

to select suitable flow rate. Press

to confirm or

press

to cancel the selection and return to previous menu.

Knob

Knob

Knob

Knob

Exit Key

Press

Knob

Pause Time: Time Interval between every dispense process.

Press and turn

to highlight pause time line. Press

to enter next

inter face. Turn

to select pause time. Press

to confirm or press

to cancel the selection and return to previous menu.

Knob

Knob

Knob

Knob

Exit

Key

Calibration

Pump should be calibrated when flow rate or dispense volume is beyond the
tolerance. Flow rate and dispensing volume can be calibrated separately in
different application.

Flow Rate Calibration

Press and turn

to highlight calibration line. Press

to enter next

inter face. Turn

to select Flowrate. Press

to highlight test time line.

Press

again to highlight the value of time. Turn

to select test time

(its range is from 0.5 to 30 minutes). Press

to confirm or press

to cancel the selection and return to previous menu.

Press Star t/Stop Key, the LCD displays running time and actual volume.
Running time counts down.

Turn

to input actual tested volume. Press

to confirm. The

calibration operation can be repeated.

Knob

Knob

Knob

Knob

Knob

Knob

Knob

Exit Key

Knob

Knob

Turn Knob to input
Actual Volume

If the actual volume is known in advance, enter the calibration inter face and
input the actual volume directly.

Turn

to input

Actual Volume

Knob

Dispensing Volume Calibration

The parameters can not be changed under dispensing volume calibration
operation. The defaulted parameters are the parameters set under Dispensing
Setting.

Pump stops under dispense mode, press and turn

to highlight calibration

line. Press

to enter next inter face. Turn

to select Dispensing. Press

to enter next inter face.

Press

Key, the LCD displays running time and actual volume.

Running time counts down.

Turn

to input actual tested volume. Press

to confirm.

Knob

Knob

Start/Stop

Knob

Knob

Knob

Knob

Turn

to input

actual volume

Knob

Turn

to input

actual volume

Knob

BT100-1F OPERATING MANUAL

BT100-1F OPERATING MANUAL

Содержание BT100-1F

Страница 1: ...Longer Precision Pump Co Ltd ADD Building A Chuangye Center Baoding National High Tech Industrial Development Zone Baoding Hebei China 071051 TEL 86 312 3110087 3138553 FAX 86 312 3168553 E mail longe...

Страница 2: ...Specification 1 2 2 3 3 3 4 5 5 6 6 6 6 7 7 8 8 8 8 9 9 9 9 10 11 11 12 14 14 15 15 15 16 11 IMPORTANT INFORMATION Please read operation manual carefully before operation WARNINGS Tubing breakage may...

Страница 3: ...Specification 1 2 2 3 3 3 4 5 5 6 6 6 6 7 7 8 8 8 8 9 9 9 9 10 11 11 12 14 14 15 15 15 16 11 IMPORTANT INFORMATION Please read operation manual carefully before operation WARNINGS Tubing breakage may...

Страница 4: ...1 Signal Output Port P2 External Control Input Port P3 Power Socket P4 Infuse Holder Basic Operation Start Stop Key Direction Key Knob Knob Knob Knob Prime Key Prime Key Prime Key Exit Key Exit Key Po...

Страница 5: ...1 Signal Output Port P2 External Control Input Port P3 Power Socket P4 Infuse Holder Basic Operation Start Stop Key Direction Key Knob Knob Knob Knob Prime Key Prime Key Prime Key Exit Key Exit Key Po...

Страница 6: ...e flow rate b Current speed Display current speed the can increase or decrease the speed c Running direction Indicate the direction of the pump d Running status stands for pump running stands for pump...

Страница 7: ...e flow rate b Current speed Display current speed the can increase or decrease the speed c Running direction Indicate the direction of the pump d Running status stands for pump running stands for pump...

Страница 8: ...ump is running Pump head and tubing setting Press the to enter interface press again to enter submenu of then press again the pump will show all the suitable pump head turn to select needed pump head...

Страница 9: ...ump is running Pump head and tubing setting Press the to enter interface press again to enter submenu of then press again the pump will show all the suitable pump head turn to select needed pump head...

Страница 10: ...uence the filling accuracy Set Press and turn to highlight the back suction time line Press the to enter next interface turn to adjust press to confirm or press to cancel the adjustment and return to...

Страница 11: ...uence the filling accuracy Set Press and turn to highlight the back suction time line Press the to enter next interface turn to adjust press to confirm or press to cancel the adjustment and return to...

Страница 12: ...selection and return to previous menu Knob Knob Knob Knob Exit Key Calibration Pump should be calibrated when flow rate or dispense volume is beyond the tolerance Flow rate and dispensing volume can b...

Страница 13: ...selection and return to previous menu Knob Knob Knob Knob Exit Key Calibration Pump should be calibrated when flow rate or dispense volume is beyond the tolerance Flow rate and dispensing volume can b...

Страница 14: ...e working Thick wall tubing is preferred for longer service life Flow rate mode operation procedure Confirm the flow rate of the fluid which need to be pumped Press to enter interface to set pump head...

Страница 15: ...e working Thick wall tubing is preferred for longer service life Flow rate mode operation procedure Confirm the flow rate of the fluid which need to be pumped Press to enter interface to set pump head...

Страница 16: ...n the pump runs clockwise Close the pumps runs counter clockwise COM of external control input Terminal 1 Terminal 2 Terminal 3 Terminal 4 COM of external control input External control start stop inp...

Страница 17: ...n the pump runs clockwise Close the pumps runs counter clockwise COM of external control input Terminal 1 Terminal 2 Terminal 3 Terminal 4 COM of external control input External control start stop inp...

Страница 18: ...ocol Maintenance When the pump is idle we recommend you to release the tubing from pressure This helps to protect the tubing from unnecessary strain and prolongs its service life Keep rollers clean an...

Страница 19: ...ocol Maintenance When the pump is idle we recommend you to release the tubing from pressure This helps to protect the tubing from unnecessary strain and prolongs its service life Keep rollers clean an...

Страница 20: ...Longer Precision Pump Co Ltd ADD Building A Chuangye Center Baoding National High Tech Industrial Development Zone Baoding Hebei China 071051 TEL 86 312 3110087 3138553 FAX 86 312 3168553 E mail longe...

Отзывы: