9
1 0
If the actual volume is known in advance, enter the calibration inter face and
input the actual volume directly.
Disp Vol
2.02mL
Disp T.
60.5"
:
:
Disp T.
60.5
Act Vol
2.020 mL
:
:
"
Act Vol
:
Flow 2.043mL/min
:
Pause T.
:
"
1.0
Pause Time
:
"
1.1
Flowrate
Dispensing
Test T.
0.5 ’
Cal Vol
132.3 L
:
:
μ
Test T.
Cal Vol
132.3 L
:
:
μ
30
"
2.02 mL
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Flow 2.043mL/min
:
Pause T.
:
"
1.0
Flow 2.043mL/min
:
Pause T.
:
"
1.0
Test T.
0.5 ’
Cal Vol
132.3 L
:
:
μ
Test T.
0.5 ’
Cal Vol
132.3 L
:
:
μ
Test T.
Cal Vol
133.0 L
:
:
μ
30
"
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Flowrate
Dispensing
Test T.
0.5 ’
Cal Vol
132.3 L
:
:
μ
Test T.
0.5 ’
Cal Vol
132.3 L
:
:
μ
:
Act Vol
133.0 L
μ
Flowrate
Dispensing
Disp T.
60.5
Act Vol
2.040 mL
:
:
"
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Flowrate
Dispensing
Disp Vol
2.02mL
Disp T.
60.5"
:
:
Flow Rate: Adjusting the dispensing flow rate can change the dispensing time.
Press and turn
to highlight dispensing flow rate line. Press
to enter
next inter face, turn
to select suitable flow rate. Press
to confirm or
press
to cancel the selection and return to previous menu.
Knob
Knob
Knob
Knob
Exit Key
Press
Knob
Pause Time: Time Interval between every dispense process.
Press and turn
to highlight pause time line. Press
to enter next
inter face. Turn
to select pause time. Press
to confirm or press
to cancel the selection and return to previous menu.
Knob
Knob
Knob
Knob
Exit
Key
Calibration
Pump should be calibrated when flow rate or dispense volume is beyond the
tolerance. Flow rate and dispensing volume can be calibrated separately in
different application.
Flow Rate Calibration
Press and turn
to highlight calibration line. Press
to enter next
inter face. Turn
to select Flowrate. Press
to highlight test time line.
Press
again to highlight the value of time. Turn
to select test time
(its range is from 0.5 to 30 minutes). Press
to confirm or press
to cancel the selection and return to previous menu.
Press Star t/Stop Key, the LCD displays running time and actual volume.
Running time counts down.
Turn
to input actual tested volume. Press
to confirm. The
calibration operation can be repeated.
Knob
Knob
Knob
Knob
Knob
Knob
Knob
Exit Key
Knob
Knob
Turn Knob to input
Actual Volume
If the actual volume is known in advance, enter the calibration inter face and
input the actual volume directly.
Turn
to input
Actual Volume
Knob
Dispensing Volume Calibration
The parameters can not be changed under dispensing volume calibration
operation. The defaulted parameters are the parameters set under Dispensing
Setting.
Pump stops under dispense mode, press and turn
to highlight calibration
line. Press
to enter next inter face. Turn
to select Dispensing. Press
to enter next inter face.
Press
Key, the LCD displays running time and actual volume.
Running time counts down.
Turn
to input actual tested volume. Press
to confirm.
Knob
Knob
Start/Stop
Knob
Knob
Knob
Knob
Turn
to input
actual volume
Knob
Turn
to input
actual volume
Knob
BT100-1F OPERATING MANUAL
BT100-1F OPERATING MANUAL