Sensor Drift correction
If your system is normally reading 0, for example as is typical with platform scales or a
torque meter, you may find a small amount of sensor drift caused by changes in
temperature, ageing etc.
We can automatically correct for slow, long term drift, by comparing your signal to 0 every
30 seconds, and re-calibrating to remove any detected movement.
This does not affect the sensititivity of your system, and is only applied for readings less
than 8 least-significant counts.
The corrected value is stored in non-volatile memory, so that any drift trend will be re-ap-
plied after you have switched off the meter.
Lockout Switch must be OFF
Press together, briefly
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Press OK button briefly and repeat-
edly until you see
drift.0
or
drift.1
displayed.
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Press briefly to toggle
1
2
3
4
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Done!
Press DOWN or UP button to toggle
between
drift.0
or
drift.1
drift.0
= No Correction
(Default)
drift.1
= Correction is active
This feature is available in Easy and Advanced Modes
Press to accept