Maintenance
Module 8023
6-12
Relay Contact Information
Connections are made depending upon detector application, refer to your
customised wiring diagrams in your User Manual supplied with the detector.
Contacts are shown are for relays when de-energised (failsafe).
Figure 3 Relay contacts
Access to the relay connections is as access to the incoming ac protection fuse
& reject relay protection fuses
on page 11
for boards
protected by a one piece safety mesh cover. For boards with the two piece
cover only the angled terminal cover access plate requires removal.
AC POWER CONNECTION INFORMATION
The location and wiring connection is shown in
on page 12
in this section.
WARNINGS
LETHAL HAZARD - ELECTRICAL SUPPLIES.
A current of 100 mA passing
through the body for one second can kill. This can occur at voltages as low as
35V ac or 50V dc.
LETHAL HAZARD -
The control board and power supply are combined on a
single board therefore this board contains high AC/DC voltages. A grounded
mesh screen cover protects the power supply area of the main control board.
Failure to fit or refit this cover securely may expose the user to a risk of electric
shock or may result in permanent damage to the detector.
AC wiring colours will vary according to territory.
North America
Black
=
HOT
110/120v
White
=
NEUTRAL
Green/Yellow
=
EARTH
All Others
Brown
=
LIVE
200/240v
Blue
=
NEUTRAL
Green/Yellow
=
EARTH
NC C NO NC C NO
Reject relay 2 contact sets
NC C NO NC C NO
Fault relay 2 contact sets
NC C NO NC C NO
Error relay Run relay
Pin1
Содержание IQ2
Страница 92: ...Maintenance Module 8023 6 1 Frequency links Tx Inductors Balance inductors ...
Страница 101: ...Wiring Diagrams Module 10088 8 2 Figure 1 Optional External Sensor Connection Ladder Diagram ...
Страница 102: ...Spare Parts Module 11050 9 1 SPARE PARTS CONTENTS Spare Parts Lists 2 ...
Страница 104: ...Service Return Module 12005 8 1 SERVICE RETURNS FORMS CONTENTS Instructions 2 Form 3 ...