Operating instructions
ScanAfia X TK, S, ICU Version 1.5
Copyright © Lojer Oy, 2015
6
4
Safety instructions
Follow these safety instructions.
IMPORTANT!
Use only accessories approved by the manufacturer for the
specific model of bed.
IMPORTANT!
The mattress must meet the bed manufacturer's requirements
for height and width.
IMPORTANT!
Attach the rescue sheet to the adjustable bed surface. Never
attach the rescue sheet to the bed frame or through frame
components, because it may prevent the adjustment of the
bed surface.
I
MPORTANT!
Deliberate or unintentional use of the bed's hand-held control
must be prevented by placing it out of reach of restless
patients and children.
IMPORTANT!
Release the pushbutton of the hand-held control to stop any
movement of the bed. In case of malfunction, press the
pushbutton of a reverse movement in the hand-held control,
nurse's locking board or nurse's control board to stop a
movement for the as long as the button is pushed. The
movement can also be stopped by locking the function in the
nurse's locking board or control board.
IMPORTANT!
Do not press two or more pushbuttons of the hand-held
control at the same time.
IMPORTANT!
The electric components under the bed surface heat up when
electric operations are used. Avoid unnecessary contact with
the electric components under the bed surface. Contact with
the components is only allowed during maintenance and
cleaning procedures.
IMPORTANT!
Allow the electric components under the bed surface to cool
off before carrying out their maintenance or cleaning
procedures. Maximum duration of continuous contact between
these components and bare skin is 1 minute.
IMPORTANT!
The rails of the bed must be freely adjustable up and down so
that the rail is not supported or wedged in the upper position
by an external obstacle (mattress, bed linen, mattress air
tubes or other object) without engaging the rail's own locking
mechanism.