LogTag
®
-
20 Day Pharmaceutical Temperature Indicator with display
© Copyright 2013, LogTag Recorders Limited
page 6 of 17
Operation
Starting the Indicator
READY State
A short press of either button will activate the
display.
If
READY
is displayed, the Indicator can be started.
The Battery condition is also displayed as
-
OK (
)
-
Low(
)
After 30 seconds the Indicator will re-enter a
hibernated state and the display will switch off.
To start the Indicator press and hold the START
button for 5 seconds.
The
MONITORING
indicator will flash.
When
MONITORING
is permanently lit, release the
START button within 2 seconds.
START DELAY State
The indicator has now started its 60 minute
START
DELAY
countdown. It should be immediately
placed in the load to be monitored so that the
temperature is stable and within specification once
the monitoring begins.