Appendix 1 - Battery Replacement
TREL30-16 User Guide
42
Appendix 1 - Battery Replacement
Prepare the Logger
l
Download the unit in LogTag
®
Analyzer to save any stored data.
l
Hibernate the unit.
Failing to hibernate the logger may
result in serious memory
corruption. As a result you may no
longer be able to use this product.
Remove the old battery
l
Carefully remove the rear label
covering the battery door using a
small-bladed screwdriver or knife:
l
Turn the battery door anti-clockwise
and remove it from the case:
You can use the
new battery for
this, as it will fit
into the slot
l
Remove the battery using a small-
bladed screwdriver, as shown:
Insert the new battery
Only use batteries from a reputable
manufacturer! Check the
temperature range of the battery
and make sure it covers the range of
the TREL30-16.
l
Place a new CR2032 cell into the
battery compartment as shown; the
right-hand edge of the battery is
inserted first against the contact:
l
Press firmly down on the left-hand
side of the battery to click the battery
into place:
l
Replace the battery door; turn
clockwise to lock it into place (using
the old battery).
l
Cover the battery door with either a
new label (Part # 100-000502) or
reuse the old one.
l
Re-configure the logger with LogTag
®
Analyzer.
Please recycle or dispose of the old
battery according to your local
environmental regulations.