3 - Setup instructions Edan EC12 Express
3 - Setup instructions Edan EC12 Express
3.1 - Cable selection
3.1 - Cable selection
You will need the RS-232 cable
(part no. MS1R- 107242,
provided by EDAN) Express to
establish communication
between the Lode Treadmill
Valiant and SE-12 Express.
Connect the RS232 cable to the
SE-12 (picture).
Connect the cable to the
ergometer as described in
chapter "Hardware setup".
3.2 - Software Setup
3.2 - Software Setup
1. Switch on SE-12 Express and
start program.
2. Click on the General Info key
on the System Setup interface.
3. Select Treadmill in the Device
Type pull-down list.
4. Select TMX425 in the Device Model pull-down list.
5. Press the OK key.
The Edan ECG is now ready for communication with the Treadmill.
4/22