13.2 - Immunity test
The Lode equipment is intended for use in the electromagnetic environment specified below. It is the
responsibility of the customer or user to ensure that the Lode equipment is used in such an
environment.
Immunity Test:
Immunity Test: Electrostatic discharge (ESD) EN 61000-4-2
IEC 60601-Test Level:
IEC 60601-Test Level: ± 6 kV contact
± 8 kV air
Compliance Level:
Compliance Level: ± 6 kV contact
± 8 kV air
Electromagnetic Environment - Guidance:
Electromagnetic Environment - Guidance: Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at least 30%.
Immunity Test:
Immunity Test: Electrical fast transient/burst EN 61000-4-4
IEC 60601-Test Level:
IEC 60601-Test Level: Not applicable
Compliance Level:
Compliance Level: Not applicable
Electromagnetic Environment - Guidance:
Electromagnetic Environment - Guidance: Conducted only on ports interfacing with cables whose total
length, according to the manufacturer’s functional specification, may exceed 3 meters
Immunity Test:
Immunity Test: Surge EN 61000-4-5
IEC 60601-Test Level:
IEC 60601-Test Level: Not applicable
Compliance Level:
Compliance Level: Not applicable
Electromagnetic Environment - Guidance:
Electromagnetic Environment - Guidance: Conducted only on ports interfacing with cables whose total
length, according to the manufacturer’s functional specification, may exceed 3 meters
Immunity Test:
Immunity Test: Voltage dips, short interruptions and voltage variations on power supply input lines EN
61000-4-11
IEC 60601-Test Level:
IEC 60601-Test Level: Not applicable
Compliance Level:
Compliance Level: Not applicable
Electromagnetic Environment - Guidance:
Electromagnetic Environment - Guidance: Equipment for professional use with a total rated power
greater than 1 kW and is excluded from this requirement
Immunity Test:
Immunity Test: Power frequency (50/60Hz) magnetic field EN 61000-4-8
IEC 60601-Test Level:
IEC 60601-Test Level: 3 A/m
Compliance Level:
Compliance Level: 3 A/m
Electromagnetic Environment - Guidance:
Electromagnetic Environment - Guidance: Power frequency magnetic fields should be at levels
characteristics of a typical location in a typical commercial or hospital environment.
NOTE:
NOTE:
Ut
is the AC mains voltage prior to application of the test level.
18/24
M960900 / VERSION: 1.0 - 2014-09-24