73
D.3
Abort codes BP system
Prior.
Error message
Display
Possible error cause
1
Α
CHECK ECG
SIGNAL
≅
0 / 0 mmHg
!
ECG trigger not connected.
!
ECG trigger properties not correct.
!
ECG trigger polarity not correct.
!
QRS pulse width
smaller than 10 ms.
2
Α
CUFF
OVERPRESSURE
≅
0 / 0 mmHg
!
Strong arm movements during the blood
pressure measurement.
!
Air hose is bent or blocked.
3
Α
AIR LEAK
≅
0 / 0 mmHg
!
Cuff is loose or not connected.
!
Air hose or air-bubble leak.
!
Air hose not connected.
4
Α
SERVICE
REQUIRED
≅
BP
value
!
Fatal error ...Get in touch with the company
Lode.
5
Α
CHECK ECG/
MICROPHONE
≅
BP
value
!
Poor patient hook-up (irregular ECG-signal)
or poor trigger signal.
AND
!
Abnormal
microphone-signal
or damaged
microphone.
6
Α
CHECK
MICROPHONE
≅
BP
value
!
Wrong Microphone placement or defect
microphone.
7
Α
CHECK ECG/
ARTIFACT
≅
BP
value
!
ECG trigger signal failed during blood
pressure measurement.
Bad or irregular ECG-signal during the blood
pressure measurement.
AND
!
Excessive movements of the patient during the
blood pressure measurement.
8
Α
ARTIFACT/ARM
MOVEMENT
≅
BP
value
!
Excessive movements of the patient during the
blood pressure measurement.
9
Α
CUFF INFLATION
TOO LOW
≅
BP
value
!
Maximum cuff pressure adjusted too low.
!
Inflation
pressure
adjusted too low. Repeat
measurement.
10
None
BP
value
D.4
Cleaning and disinfection of the cuff