Loctite
®
Zeta
7021 Spot Radiometer
Part Number 98048
EQUIPMENT
Operation Manual
985706 Zeta 7021 Rev A 4/17/02 4:15 PM Page 1
Страница 1: ...Loctite Zeta 7021 Spot Radiometer Part Number 98048 EQUIPMENT Operation Manual 985706 Zeta 7021 Rev A 4 17 02 4 15 PM Page 1 ...
Страница 2: ... 2 1 Operation 6 3 Technical Data 7 8 3 1 Specifications 7 3 2 Calibration 8 4 Installation 9 4 1 Unpacking and Inspection 9 4 2 Installation and Start up 9 5 Operations 10 5 1 Controls and Operation 10 6 Troubleshooting 11 7 Care and Maintenance 12 7 1 Maintenance 12 8 Spare Parts List 12 9 Warranty 13 985706 Zeta 7021 Rev A 4 17 02 4 15 PM Page 2 ...
Страница 3: ...erty is not damaged Notice Gives recommendations for better handling of the unit during operation or adjustment as well as for service activities 1 2 For Your Safety IMPORTANT DO NOT OPERATE THE UNIT BEFORE READING THIS SECTION The Loctite UV Spot Radiometer will be used to measure the output of UV Wand Products Always assure that the Wand is held securely in the Wand Sensor Head and inserted to t...
Страница 4: ...tite Safety Glasses Item 97210 2 Potentially exposed skin should be covered Long sleeves and protective gloves should be worn when measuring UV Wand intensity 3 Never look directly at the output from the UV Wand 4 Make Sure that the Radiometer reading is Zero before removing the UV Wand from the Sensor Head When in doubt about UV safety consult a Loctite technical specialist at 1 800 LOCTITE 562 8...
Страница 5: ...easures the UV Intensity emitted from a 3 mm or a 5 mm UV wand The Intensity is measured from 280 400 nm UV A B The Radiometer is traceable to the National Institute of Standards and Technology NIST 3 985706 Zeta 7021 Rev A 4 17 02 4 15 PM Page 5 ...
Страница 6: ...e Resolution 280 400nm UV A B 370nm 5 to 10 max GaAsP Photodiode 9V Battery 0 19 99W cm 10mW cm 2 2 O I 0 00 Sensor Connector Panel Magnet Digital Panel Meter Display On Off Rocker Switch Front View Side View Back View Top Cover Battery Door Figure 1 985706 Zeta 7021 Rev A 4 17 02 4 15 PM Page 6 ...
Страница 7: ...2 Description continued Sensor Head Assembly Sensor Head Cap Sensor Head Magnet 3mm and 5mm Light Guide Sensor Plug Filter Glass Figure 2 5 985706 Zeta 7021 Rev A 4 17 02 4 15 PM Page 7 ...
Страница 8: ... production operations The UV Spot Radiometer can be used to verify that the output from the wand system under test is proper for the curing process When making measurements the wand should be held as straight as possible and the ends should be clean 985706 Zeta 7021 Rev A 4 17 02 4 15 PM Page 8 ...
Страница 9: ...e Battery Not Included Dimensions L x W x D inches 3 0 x 2 5 x 1 05 Total Weight 5 6 oz Spectral Range 280 400 nm UV A B Range 0 19 99 Watts cm squared Resolution 10 mw cm squared Accuracy 5 typical 10 max Traceable to NIST Wand Choices 3 mm and 5 mm 3 Technical Data 985706 Zeta 7021 Rev A 4 17 02 4 15 PM Page 9 ...
Страница 10: ...ional Institute of Standards and Technology NIST primary physical standard for UV photodiodes The recommended calibration frequency is every 12 months Contact Loctite Corporation at 1 800 571 5174 for re calibration services 3 Technical Data continued 985706 Zeta 7021 Rev A 4 17 02 4 15 PM Page 10 ...
Страница 11: ...any missing parts promptly to the Loctite customer service department at 1 800 LOCTITE 562 8483 4 2 Installation and Start up The Loctite ZETA 7021 UV Spot Radiometer will arrive completely assembled All that is required is to plug in the Sensor Head and install a 9 Volt Alkaline Battery Not included with unit Battery is installed by sliding off rear cover see section 2 Turn on switch and you are ...
Страница 12: ...s easy to use It requires only that a 9 volt Alkaline Battery be installed and the power switched on to operate The panel meter displays the power reading in Watts per centimeter squared 5 Operations 985706 Zeta 7021 Rev A 4 17 02 4 15 PM Page 12 ...
Страница 13: ... functional 2 Meter Reads only 1 Make sure that the Sensor Head is plugged zero or intermittent in properly 3 UV Output is low 1 Make sure wand is seated in the bottom Radiometer of the Sensor head Reading 2 Unscrew Sensor Head Cap and clean filter glass located under adapter 3 Make sure wand ends are clean 985706 Zeta 7021 Rev A 4 17 02 4 15 PM Page 13 ...
Страница 14: ...e glass can be accessed by unscrewing the Sensor Head Cap see figure section 2 Cleaning can be performed with a cotton swab and isopropyl alcohol Low Battery When battery is low a small battery symbol will appear in the upper left hand corner of the display indicating the 9 volt Alkaline battery needs to be replaced 7 Care and Maintenance Sensor Head 985705 Manual 985706 8 Spare Parts List 985706 ...
Страница 15: ...use of products parts or attachments which are not manufactured by Loctite No Products shall be returned to Loctite for any reason without prior written approval from Loctite Products shall be returned freight prepaid in accordance with instructions from Loctite NO WARRANTY IS EXTENDED TO ANY EQUIPMENT WHICH HAS BEEN ALTERED MISUSED NEGLECTED OR DAMAGED BY ACCIDENT OR IF THE SYSTEM IS USED TO DISP...
Страница 16: ...g Rocky Hill CT 06067 3910 Loctite Automotive Technology Center 2455 Featherstone Road Auburn Hills Michigan 48326 Loctite Canada Inc 2225 Meadowpine Boulevard Mississauga Ontario L5N 7P2 Loctite Brazil Av Prof Vernon Krieble 91 06690 11 Itapevi São Paulo Brazil Loctite Company de México S A de C V Calzada de la Viga s n Fracc Los Laureles Loc Tulpetlac C P 55090 Ecatepac de Morelos Edo de México ...