background image

E-mail support 

[email protected]

 

LMI Corporation

Phone support 

(810) 714-5811 

 

Form:  CA 068 

August 01, 2005 

R:\Quality\Calibration Instructions\CA 097 200s or sb to LMI 585 Plus.doc 

Page 4 of  4 

 

 
 
 
 
 
 
 
 

NOTES: 
1)

 If you are using the 200-SB with the 235/236 seal gap blocks, you will want to CAL MASTER on 

the LO step of the Master Block and put in the proper offset in the ADVANCED SETUP>PORT 
CONFIG>Port1a menu.  The offset should be set to the 235/236 minimum gap check (10.0mm typ.). 

2)

 A 200-SF Probe has 22.0mm of travel so the CALSPAN needs to be changed from the default 

10.00 to 22.00mm.  Go to the ADVANCED SETUP>PORT CONFIG>Port1a menu to adjust the 
CALSPAN to 22.00mm.  The 210 Master Block will also be replaced with the 214 Master Block 
with a 54.0mm Master. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

¾

 

To reverse the signs, go to the main menu.  Go to Advanced setups and hit <SELECT>.  Go to Ports 
Polarity.  Press <SELECT>.  Choose the port you are using and press <SELECT>…the port readings 
will change to “Inverted”

 

Отзывы: