Electromagnetic Compatibility Information

Recommended separation distances between portable and mobile  RF communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT 

or SYSTEM that are not LIFE-SUPPORTING

Recommended separation distances between portable and mobile RF communications equipment and the BM52 blood 

pressure monitor
The BM52 blood pressure monitor is intended for use in an electromagnetic environment in which radiated RF disturbances 

are controlled. The customer or the user of the BM52 blood pressure monitor can help prevent electromagnetic interference by 

maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the BM52 

blood pressure monitor as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output 

power of transmitter

(W)

Separation distance according to frequency of transmitter

(m)

150 kHz to 80 MHz

d = 

 

80 MHz to 800 MHz

d = 

 

800 MHz to 2.5 GHz

d = 

 

0.01

0.117

0.117

0.234

0.1

0.370

0.370

0.740

1

1.170

1.170

2.340

10

3.700

3.700

7.400

100

11.7

11.7

23.4

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) 

can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power 

rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1  At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2   These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

                reflection from structures, objects and people.

Beurer GmbH, Söflinger Str. 218, 89077 Ulm, Germany  

www.beurer-medical.de   www.beurer.com

751.583-0316 

       V7 

160404

26

Содержание BM52

Страница 1: ...User Manual BM52 Speaking Blood Pressure Monitor ...

Страница 2: ...8 19 20 21 22 23 26 Thank you for purchasing the LloydsPharmacy Speaking Blood Pressure Monitor Please read these instructions for use carefully and keep them for later use be sure to make them accessible to other users The upper arm blood pressure monitor is used for non invasive measurement and monitoring of adults arterial blood pressure You can use it to measure your blood pressure quickly and...

Страница 3: ...y are not a substitute for a medical examination Discuss the measurements with your doctor and never base any medical decisions on them e g medicines and their administration Do not use the blood pressure monitor on newborns or patients with pre eclampsia We recommend consulting a doctor before using the blood pressure monitor during pregnancy This device is not intended for use by people includin...

Страница 4: ...information otherwise erroneous results will be recorded If the instrument is not to be used for any length of time we recommend removing the batteries Notes on handling batteries If your skin or eyes come into contact with battery fluid flush out the affected areas with water and seek medical assistance Choking hazard Small children may swallow and choke on batteries Store the batteries out of th...

Страница 5: ...cuff connector left hand side Information on the Display Inflate release air Systolic pressure Diastolic pressure Pulse value Memory display average value AVG morning AM evening PM Cardiac arrhythmia symbol Time and date Battery replacement symbol Multi user memory Classification of measurements Memory space number 6 7 Memory button ...

Страница 6: ...d The device will announce Battery weak Please replace all batteries Once the batteries have been removed from the unit the date time and language must be set again 4 x AA LR6 1 5 V Setting up your Blood Pressure Monitor Setting up your Blood Pressure Monitor Set date time and language This menu allows you to set the following functions one after another Date Time Language Volume It is essential t...

Страница 7: ...e function buttons Vo3 loud Vo2 medium Vo1 quiet Confirm your selection with the Memory button How to use your Blood Pressure Monitor If the measurement is performed on the right upper arm the line should be located on the inside of your elbow Ensure that your arm is not pressing on the line Fit the cuff on your arm Blood circulation in the arm should not be restricted by tight clothing or other o...

Страница 8: ...e that the cuff is on a level with your heart To carry out a blood pressure measurement make sure you are sitting comfortably with your arms and back leaning on something Do not cross your legs Place your feet flat on the ground In order not to distort the result it is important to keep still during the measurement and do not talk Selecting memory Selecting memory Voice output when voice function ...

Страница 9: ...a pulse is detected the symbol flashes Measuring can be cancelled at any time by pressing the Start stop button Systolic pressure diastolic pressure and pulse measurements are displayed Systole mmHg Diastole mmHg Heart rate beats per minute According to the WHO guidelines your blood pressure is optimal normal high to normal shows slight hypertension shows medium hypertension shows strong hypertens...

Страница 10: ...BP fall into different categories the higher value should be taken for classification e g if your BP is 141 83 you should read the advice in the Hypertension category not the normal category Your BP should be re assessed professionally at least once every 5 years unless you have high BP in which case it should be re assessed at least once a year speak to your LloydsPharmacy pharmacist or healthcar...

Страница 11: ... will be displayed morning 5 a m 9 a m display If you press the Memory button again the average value of the evening measurements for the last 7 days will be displayed evening 6 p m 8 p m display Average count Systole mmHg Diastole mmHg Heart rate beats per minute According to the WHO guidelines your blood pressure is Saving displaying and deleting measurements Individual measured values If you co...

Страница 12: ... not be bent sharply Technical specifications Model no BM 52 Measurement method Oscillometric non invasive blood pressure measurement on the upper arm Measurement range Cuff pressure 0 300 mmHg systolic 50 250 mmHg diastolic 30 200 mmHg Pulse 30 180 beats minute Display accuracy Systolic 3 mmHg diastolic 3 mmHg pulse 5 of the value shown Measurement inaccuracy Max permissible standard deviation ac...

Страница 13: ...ood pressure monitor is intended for use in the electromagnetic environment specified below The customer of the user of the BM52 blood pressure monitor should assure that it is used in such and environment Emission test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The BM52 blood pressure monitor uses RF energy only for its internal function Therefore its RF emissio...

Страница 14: ...rom fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Electromagnetic Compatibility Information Guidance and manufacture s declaration electromagnetic immunity NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NO...

Страница 15: ...of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 80 MHz to 800 MHz d 800 MHz to 2 5 GHz d 0 01 0 117 0 117 0 234 0 1 0 370 0 370 0 740 1 1 170 1 170 2 340 10 3 700 3 700 7 400 100 11 7 11 7 23 4 For transmitters rated at a maximum output power not listed above the recommended separation dista...

Отзывы: