If the LIFT CLOSE is active and the HEAD OPEN button is pressed, then the 
BACK section will begin to OPEN. Pressing the HEAD CLOSE or BACK  CLOSE 
button will stop the BACKmotor.

If the LIFT CLOSE is active and the BACK CLOSE button is pressed, then the 
BACK section will begin to CLOSE. Pressing the HEAD OPEN or BACK  OPEN 
button will stop the BACKmotor.

If the LIFT CLOSE is active and the HEAD CLOSE button is pressed, then  the 
BACK section will begin to CLOSE. Pressing the HEAD OPEN or BACK  OPEN 
button will stop the BACKmotor.

Error:
BACK or LEG/LIFT
motor doesnot  

function

If the BACK or LEG/LIFT motor(s) do(es) not function, the motor connection(s)  

could have become disconnected.

Please follow these steps to attempt to correct the error:

1)

Unplug the piece of furniture from the wall outlet

2)

Wait at least 30 seconds

3)

Lift the back panel to reveal zipper pouch

4)

Unzip pouch to reveal the Junction

5)

Locate the BACK or LEG/LIFT Motor Cord (shown below)

6)

Ensure that there are no frayed, exposed, or otherwise damaged wires.  Do 
not connect the motor if the cord is damaged in anyway.

7)

Ensure that the BACK or LEG/LIFT Motor Cord (Motor Connection) is  
fully seated and attached to the motor

8)

Ensure that there is nothing blocking the path of the motor orframe

9)

Locate the BACK or LIFT/LEG Motor Cord (Junction Connection) (shown  
below)

10) Ensure that there are no frayed, exposed, or otherwise damaged wires.  Do 

not connect the motor if the cord is damaged in anyway.

11) Ensure that the HEAD Motor Cord (Junction Connection) is fully seated  

and attached to the Junction

12) Zip the pouch and reset the back panel
13) If the motor still does not function, contact your furniture provider so  

they may replace or repair the faulty part

BACK or LIFT/LEG Motor Cord (Motor Connection)

BACK or LIFT/LEG Motor Cord (Junction Connection)

Hand Wand Manual

Содержание 1555 LIFT CHAIR

Страница 1: ...1555 LIFT CHAIR USER MANUAL TroubleshootingGuide This guide is to help with a non functioningpower recliner...

Страница 2: ...y the battery is mountedat the top of transformer and couldprovide the power Y cable w control box splitter extensioncord connecting the power transformer to multiple motors hand wand heating pad Back...

Страница 3: ...securely plugged in 2 Broken hand wand 3 Broken power transformer especially when ALL motors stop working on one unit 4 Power cords damaged 5 Motor failure NOTE In some cases a button on the switch m...

Страница 4: ...piece of furniture indicatedby the button icon The LEG buttons will open and close the leg rest position of the piece of furniture indicated by the button icon The HEAD buttons will open and close th...

Страница 5: ...nofrayed exposed orotherwisedamaged wires Do not connect themotor if thecord is damagedin anyway 11 Ensure thattheHEAD Motor Cord Junction Connection is fully seated and attachedto the Junction 12 Zi...

Страница 6: ...ed exposed orotherwisedamaged wires Do not connect themotor if thecord is damagedin anyway 11 Ensure thattheHEAD Motor Cord Junction Connection is fully seated and attachedto the Junction 12 Zip the p...

Страница 7: ...the pieceoffurnitureis fully lowered willcausetheLEG section to begin opening Please take cautionwhen pressing theLIFTor LEG buttons wheneither section is fully opened orclosed When the LIFT function...

Страница 8: ...s 3 Lift the back panelto revealzipper pouch 4 Unzip pouch toreveal the Junction 5 Locate theBACK orLEG LIFTMotor Cord shownbelow 6 Ensure thatthereare nofrayed exposed orotherwisedamaged wires Do not...

Страница 9: ...EAT button will be lit a dim red when in theWARMmode When heat is not activetheHEATbutton willbelitblue The heat functionwill be activewhilethe HEATbuttonis illuminated To de activateheat press theHEA...

Страница 10: ...elto revealzipper pouch 4 Unzip pouch toreveal the Junction 5 Locate theHeatCord shown below 6 Ensure thatthereare nofrayed exposed orotherwisedamaged wires Do not connectthe motor if the cordis damag...

Страница 11: ...it when idle Ifthe Hand Controldoes not correctly control themotionfeaturesofthepiece of furnitureor is not litwhileidle thesystem could have becomedisconnected fromthe wall power supply or the Hand C...

Страница 12: ...Power Lift Recliner PARTS...

Страница 13: ...Y cord with control box 1 Check power cord s integrity and connectionto transformer 2 If powercord s undamaged and connection secured replace powertransformer Yes No Power Recliner Troubleshooting For...

Отзывы: