21x100 mW
10:00 - 0,0 J/LD
100mW 30 J/LD
CW
The selected value for the therapy time is confirmed and stored by pressing the
SET key again. You can now start using the laser (
●
section 7: *Using the laser*).
Please note that the settings for the parameters are retained when the unit is
turned off and on again.
7. Using the laser
The
laser application can only be activated, if the initial setting (display of the
parameters) is shown on the screen.
Pressing the LASER key (fig. 1/4) activates the laser after a 2-sec. warm-up pe
-
riod. The LED (fig. 1/6) lights up green and the display changes to the following
screen:
The display (page 5, fig. 1/5) shows the remaining therapy time and the emitted
energy dose rate in Joule. The end of the first line displays the mode selected. The
selected therapy time now counts backwards and the energy dose rate changes
in line with the selected laser power and the elapsed therapy time, according to
the formula “Energy dose rate (Joule) = Laser power (mW) x Therpay time (sec.)”.
An audible signal sounds while the laser output is active.
Treatment ends automatically after the selected therapy time has elapsed or
when the selected energy dose has been achieved. The end of the therapy is
indicated by a short triple sound signal.
Treatment can be terminated manually at any time by pressing the LASER key
(page 4, fig. 1/4).
Once the therapy has ended, the screen changes back to the “Initial settings”
display.
page 14 of 29