background image

1

Technical description Nightwatch for Livanova

Содержание NightWatch Livanova

Страница 1: ...1 Technical description Nightwatch for Livanova ...

Страница 2: ......

Страница 3: ...arm module and the alarm unit 8 a Arm module and alarm unit are similar 8 b Specifics arm module 9 c Specifics alarm unit 9 Putting the NightWatch into use 11 a Customizing the bracelet 11 b First activation and charging arm module 13 c Putting the alarm unit into use 14 d Daily use 14 e Connecting arm module and alarm unit 15 ...

Страница 4: ...4 ...

Страница 5: ... Livassured NightWatch kit In the Livassured NightWatch kit the following parts are included 1 Alarm unit 2 Power supply for alarm unit black 3 Arm module 4 Charger for arm module white 5 Strap 6 Strap clips 5 6 3 4 1 2 ...

Страница 6: ... of most adults the NightWatch gives an alarm in regard to many clinically urgent seizures Furthermore when the alarm occurs the tests show that up until now in more than half of the events indeed there is an urgent attack However it may occur that the NightWatch functions less well In most of those instances too many false alarms take place Causes are too much mobility sleep that is too light and...

Страница 7: ...ition This has to be done one time only while putting the NightWatch to use activating the arm module This activation is confirmed by the two heart rate LEDs lighting up After this one may connect the arm module to the charger The white side of the arm module contains one red and one green LED which give an array of signals Alarm unit The white square box is the alarm unit that delivers both light...

Страница 8: ... arm module pick up the heartbeat sooner keep the arm still More than 30 seconds no detection of heartbeat and movement This happens for instance when the bracelet is not on the arm of the wearer anymore or when the arm module has not been connected to the charger Go to the wearer and put the bracelet back on or connect the arm module to the charger The signal stops by itself when a heartbeat is d...

Страница 9: ...charged battery power is ok Green LED blinks Off the charger Arm module in vertical position and or normal movement or On the charger Arm module is charging c Specifics alarm unit Orange LED burns continuously The alarm unit is connected to electricity grid One green LED burns continuously and a second green LED blinks The arm module is connected to the charger and still is charging Two green LEDs...

Страница 10: ...nd NOTE the technical alarm may be made softer but cannot be shut off totally Use the buttons to adjust the volume of both the seizure alarm and the technical alarm Press on button a or b You will hear the sound of the alarm once Press another time on a louder or b softer to adjust the sound Repeat this to your liking NOTE do not turn the sound down so much that you risk not hearing it in case of ...

Страница 11: ...eter of the arm of the wearer above the biceps while the arm is bent Do not stretch the strap Mark with the pencil the point where the overlap starts Cut the strap at the marking point Attach the two clips on the strap as depicted below 1 Measure the contour of your forearm without stretching the strap Bend your arm when you measure Bend your arm when you m 2 3 4 Attach the 2 clips on the strap At...

Страница 12: ... problem if after the night the arm module gives a slight imprint on the arm of the wearer as long as this imprint vanishes after 5 to 10 minutes NOTE the strap is meant to keep sticking out a little The strap is very soft and the wearer will not feel it This sticking out is needed to be able to make the bracelet a lit tle tighter on a later moment might this become necessary 3 4 Attach the 2 clip...

Страница 13: ...otally charged This process may last up to 12 hours Connect the charger to the arm module and put the plug in the socket The arm module must always be on the charger when not in use It cannot be over charged As soon as you put the arm module on the charger the heartbeat LEDs on the dark grey side stop burning and the green LED on the white side starts to blink This indicates that charging is happe...

Страница 14: ... use Roll up the sleeve of the wearer Take the arm module from the charger The two bright green LEDs on the dark grey side start to burn The red LED on the white side of the arm module starts to blink indicating that no heartbeat is detected yet Put the bracelet on the upper arm of the wearer Put it on the front of the biceps so the wearer will not lie on it when he she is lying on one side Wait u...

Страница 15: ...OFF and ON again then throw the arm module three times 1 meter in the air and catch it The LED on the arm module should turn orange now If this does not happen please try this step again If the LED turns orange perform step 2 2 2 2 If the LED on the arm unit is orange connect the power supply of the alarm unit and wait for it to start up 30 seconds After startup hold both buttons of the alarm unit...

Страница 16: ...www livassured nl ...