54 of 92 — 26-0008-1100/5
(Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
7. U
SER
P
REFERENCES
S
CREEN
__________
Eight functions can be accessed from the USER PREFERENCES Screen
(see Figure 34). This screen can be accessed via the MAIN MENU screen.
Figure 34.
USER PREFERENCES Screen
7.1.
Set Timeout
By tapping the drop-down menu to the right of
Set the activity timeout for
interrogated data (minutes)
, the timeout function can be set for intervals
from 1 minute to 90 minutes. Turning the timer to OFF is also available. If
the handheld computer is idle for the amount of time indicated in the
button, the Pulse Generator must be reinterrogated before performing any
programming.
Tap
Save Settings
after making changes to this button.
Note
programming
(applicable to Model
100, 101, and 102 Pulse
Generators ONLY)” on
page 26.
This function serves as a safety device to help prevent the possibility of
accidentally programming a patient’s Pulse Generator to the settings from
a previous database inquiry or programming event. This is only applicable
to Model 100, 101, and 102 Pulse Generators.