INFORMATION ABOUT ERRORS
Indication
Likely cause
Methods of correction
The arm cuff is applied
incorrectly or the air tube
T-piece is inserted not tightly
enough.
Measurements cannot be made
due to hand movement or
talking during measurements.
The cuff has not been pumped
up to the required pressure.
At such disorders as deep
vascular sclerosis, weak pulse
wave and also in patients with
the prominent distortions of
cardiac rhythm it may be diffi
cult
to measure the blood pressure
accurately.
Be sure that the arm cuff is
applied correctly and the plug
is inserted tightly. Repeat the
whole measurement proce-
dure.
Repeat the measurement fol-
lowing strictly the recommen-
dations of this Manual.
Repeat the measurement after
pumping the cuff up 30-40
mmHg above the expected
systolic pressure.
In such cases consult a certifi ed
doctor about application of the
electronic Device.
Batteries are discharged.
Replace the batteries for new
ones.
CARE, STORAGE, REPAIR AND DISPOSAL
1. Keep this Device from exposure to higher humidity, direct sunlight, shocks,
vibration. THIS Device IS NOT WATERTIGHT!
2. Do not keep and use this Device near heating installations and open fi re.
3. Remove the batteries if the Device will be unused for a long time. Battery
leaking may damage the Device. KEEP BATTERIES OUT OF REACH OF
CHILDREN!
4. Keep the Device clean and protect it from dust. Use the dry soft cloth to
clean the Device.
5. Keep the Device and its components away from water, solvents, spirit and petrol.
6. Protect the arm cuff from contacting on sharp things; do not stretch or fold
tightly the arm cuff .