CAUTION 

  Limit the use of the device to 20 minutes at a time, and wait 40 minutes before using the device again. 

  Provide close supervision when this device is used by, on, or near infants, children or compro-

mised individuals.

  Do not insert any object into the compressor. 

  Make sure that the air 

fi

 lter is clean. If the air 

fi

 lter has changed color or has not been used for 

60 days, replace the 

fi

 lter. 

  Make sure that the nebulizer kit is correctly assembled, the air 

fi

 lter is properly installed, and the 

air tube is correctly connected to the compressor and the nebulizer kit. Air may leak from the air tube 
during use if not securely connected. 

  Inspect the compressor (main unit) and the nebulizer parts each time before using the device. 

Make sure no parts are damaged, the nozzle and air tube are not blocked and the compressor 
operates normally. 

  Do not use the device if the air tube is bent.

  Do not block the air 

fi

 lter cover. 

  Do not alter the baf

fl

 e, the nozzle in the medication tank or any part of the nebulizer kit. 

  Do not add more than 10ml of medication to the medication tank. 

  Do not operate the device at temperatures greater than 40

˚

C. 

  Do not tilt the nebulizer kit so the angle of the kit is greater than 45

˚

. Medication may 

fl

 ow into the mouth. 

  Do not shake the nebulizer kit while using the device. 

  Do not subject the compressor, or any of the components to strong shocks, such as dropping 

on the 

fl

 oor. 

  This device is approved for human use only. 

  Do not disassemble or attempt to repair the device or components. 

  Use the device only for its intended use as described in the instruction manual. Do not use 

attachments not recommended by the manufacturer.

  Dispose of the device, components and optional accessories according to applicable local regu-

lations. Unlawful disposal may cause environmental pollution. 

  Make sure that the air tube is securely attached to the compressor (main unit) and nebulizing 

parts, and does not come loose. Twist the air tube slightly when inserting it into the connectors to 
avoid the tube disconnecting during use. 

RISK OF ELECTRICAL SHOCK

  Do not use the compressor (main unit) and the power cord while they are wet. 

  Do not plug or unplug the power cord into the electrical outlet with wet hands. 

  Do not immerse the compressor (main unit) in water or other liquid. 

  Do not spill water or other liquids on the compressor .These parts are not waterproof. If liquid 

spills on these parts, please unplug the power cord and wipe off the liquid with gauze or other soft 
absorbent material immediately.

  Do not use or store the device in humid locations or outdoors. Use the device within the operat-

ing temperature and humidity.

  Do not overload power outlets. Plug the device into the appropriate voltage outlet. 

  Do not use extension cords. Plug the power cord directly into the electrical outlet. 

  Unplug the power cord from the electrical outlet after using the device. Never leave this product 

Содержание LD-210C

Страница 1: ...LD 210 RUS LD KAZ ENG UKR LD LD Compressor Nebulizer LD Instruction Manual...

Страница 2: ...ENG 38 PARTS AND COMPONENTS 2 1 3 9 10 11 2 4 5 8 6 7 15 14 16 17 18 19 13 22 B 12 1 12 7 12 6 12 4 12 5 12 A 12 2 C 12 3 B 12 4...

Страница 3: ...affle mainly the lower respiratory tract Consist of two parts cone and cylinder Expendable material 12 3 Baffle C Red LD N003 Inhalation baffle mainly the upper respiratory tract Expendable material 1...

Страница 4: ...solution aerosols This Instruction Manual is designed to assist the user with safe and effective operation of the Compressor Nebulizer LD Use this Device according to the rules described in this Manu...

Страница 5: ...mponents to strong shocks such as dropping on the floor This device is approved for human use only Do not disassemble or attempt to repair the device or components Use the device only for its intended...

Страница 6: ...ot wrap the power cord around the compressor main unit The followings are maintenance and repair which can be taken by operator or which must be oper ated by manufacturer or distributor SERVICE AND MA...

Страница 7: ...affle C For a more effective impact on the lower respiratory tract baffle A yellow color which consists of two parts Fig 2 figures indicate the order of assembly Graphics of the differential particle...

Страница 8: ...total mass 1 100 60 30 10 7 5 3 0 5 10 15 particle sizes mkm to the total mass 1 100 60 30 10 7 5 3 0 5 10 15 Measured by Little Doctor International S Pte Ltd Performing the inhalation The length of...

Страница 9: ...t Clean and wash the device as described in last paragraph SAFETY INFORMATION WARRANTY The following LD product is covered by warranty for the period 36 months The warranty does not apply to the consu...

Страница 10: ...nutes on 40 minutes off Operation Temperature and humidity atmospheric pressure 10 C 35 C 15 80 860 1060hPa Transport and storage temperature and humidity atmospheric pressure 20 C 40 C 15 95 500 1060...

Страница 11: ...g use Turn the device off Wait 40 minutes before using the device again CERTIFICATION AND STATE REGISTRATION This device manufacturing is certified according to international standard ISO 13485 Device...

Страница 12: ...Little Doctor International S Pte Ltd Little Doctor International S Pte Ltd 2020 I408 2010 26 UA TR 001 WWW LITTLEDOCTOR SG...

Отзывы: