  Heat Exchanger 

Instruction Manual 

 

HX-0100C, HX-0300C Series 

 

Safety precautions 

 

Incorrect handling may cause death or injury

 

  (1)  Before connecting with the fittings, check if no damage or problems are found on 

the fittings. Connect properly and make sure that leak test is conducted before 
actual operation to prevent fluid from leaking into the atmosphere

 

(Hereinafter, 

the fluid used is referred as "gas" or "fluid").

 

  (2) 

DO NOT

 apply any fluids corrosive to

 

materials exposed to gas. Corrosion 

may cause fluid to leak into the atmosphere. Please confirm the physical 
properties of

 

fluid before using. 

 

  (3)  This device is not designed as an explosion-proof structure.   

DO NOT

 use this device in a place where explosion-proof structure are required. 

Doing so may cause fire or explosion.   

  (4)  Prepare temperature controller unit when operating Vaporizer/Heat 

Exchanger and do not set the temperature over than maximum operating 
temperature. Wring temperature setting may cause fire or destruction of the 
device. It is recommended to add abnormal overheating detector if 
necessary.   

  (5)  This device must be earthed before use. Otherwise, there is the risk of 

electric shock.   

  (6)  Thermal switch is equipped in Vaporizer/Heat Exchanger to prevent 

overheating. However, the operating temperature of thermal switch would 
vary due to operating conditions and ambient temperature.   

  (7)  Attach/remove connector and terminals, please make sure that power supply 

turning off. It may cause fire or shock hazard. 

 

Incorrect handling can cause medium or

 

slight injury or 

may cause

 

damage to, or loss of, facilities or equipment. 

 

  (1)  Observe the listed in the WARNING (above). 
  (2)  Use out-of-spec power supply will cause electric shock, fire, and 

malfunction of device.   

  (3)  This device is not designed to be waterproof. DO NOT locate this device 

outdoors or in a place where it may be splashed with water. Doing so may 
cause fire, trouble, or malfunction of the device.   

  (4)  DO NOT modify this device. Modification may result in fire or failure of 

the device.   

  (5)  A warm-up period of 60 minutes is recommended after reaching the set 

temperature. Otherwise, the output gas temperature will be low. 

  (6)  This device is a precious device, please handle it carefully. Dropping down 

or handing it carelessly will cause damage. Please use assist instrument 
while moving or setting the device. 

  (7)  The surface of device gets high temperature while heating up. Please wrap 

up the device to prevent careless touch. Please be careful to deal with the 
device while it is working because of the risk of burning from high 
temperature. Please conduct replacement after checking that the device has 
cooled down.

 

 

1. Introduction 

This manual explains basic operation of the HX-0100C, HX-0300C Series 
(Hereinafter referred as "HX"). Please read through this manual carefully to 
familiarize yourself with the features of HX. 

2. Summary 

HX is an ultra-clean high efficiency fluid heat exchanger unit employing Lintec’s 
high  efficiency  liquid  vaporization  technology  with  maximum  operating 
temperature  up  to  HX-0100C:300°C,  HX-0300C:250°C  and  maximum  heat 
exchange  rate  up  to  HX-0100C:N2  100SLM,  HX-0300C:N2  300SLM.  This 
apparatus  is  employed  over  a  wide  range  of  applications  from  semiconductor 
industries to other major manufacturing sectors. 

 

3. Features 

RoHS compliant. 
 
 
 
 
 
 
 
 

4. Specification/ Dimensions 

     
    (1) Specification

 

Product name 

Heat Exchanger 

Model 

HX-0100C 

HX-0300C 

Flow rate (N2) 

100SLM 

300SLM 

Pressure loss(N2) 

9.8kPa(100SLM) 

30kPa(300SLM) 

Withstand pressure 
(Gauge pressure) 

1 MPa(G) 

Leak integrity 

Less than 1×10

-11

Pa·m

3

/sec (He) 

Operating condition 

Continuous operation 

Operating temperature 

15 to 50ºC (Without dew condensation) 

Maximum 
operating temperature 

300 ºC Max 

250 ºC Max 

Recommended 
temperature control 
method 

PID control 

Material exposed to gas 

Stainless steel 316L 

Fitting 

Inlet: 12.7mm (1/2”) VCR male 

Outlet: 12.7mm (1/2”) VCR female 

Heater 

D05 

Outer block: 

120V 400W 
Inner block: 
120V 300W 

D06 

Outer block: 

240V 1200W   

Inner block: 
240V 900W 

     

D06 

Outer block: 

240V 500W 
Inner block: 
240V 375W 

Power source 

     

D05 

AC100 to 120V 

D06 

AC200 to 240V 

     

D06 

AC200 to 240V 

Thermocouple 

K type 2pcs 

Thermal switch 
Specification (Note1) 

350 ºC ± 9ºC OPEN 

Mounting position 

Free 

Weight 

Approx.5.5kg 

Standard accessories 

Heater/Thermal switch connector: SRCN6A16-7S (JAE) 
Thermocouple connector: CMP01-K (RKC) 

Note1)

 

In actual operation, there may be a difference between the temperature at the   

temperature control point and the temperature at which the thermal switch operates. 

 
 

(2) Dimensions 

 

 