How To handle video option with Chip DSC
4
3. Prepare the sample
3.1 General
o
Chose the right crucible for your needs (standard crucible 302 93 042 is recommended). It
shouldn’t be sealed and there shouldn’t be any reflecting surface
o
Prepare your sample and fill it in your crucible. The crucible should be covered with sample as
good as possible. The sample shouldn’t have any reflecting surface
right
wrong
4. Run a measurement
4.1 setup measurement settings
o
Insert name, weight, sampling interval an all other general settings in measurement software
o
Setup your temperature profile, with heating rate, target temperature and dwell time to your
needs
o
Setup additional features like Gas flow, Scheduler, temperature modulation aso.
4.2 setup video settings
o
Enable the video function by ticking