IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
The Otoscope is intended for use in the electromagnetic environment
specified below. The customer or the user of the Otoscope should assure
that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601
test level
3Vrms
150kHz to
80MHz
10V/m
80MHz to
2.7GHz
Not
applicable
(Battery
operated
device)
10V/m
Compliance
level
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Otoscope, including
cables, than the recommended
separation distance calculated
from the equation application to
the frequency of the transmitter.
Recommended separation
distance
Where p is the maximum output
power rating of the transmitter in
watts(W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meter (m)
Field strengths form fixed RF
transmitters, as determined by
an electromagnetic site survey,
a should be less than the
compliance level in each
frequency range. b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Manufacturer’s declaration – electromagnetic
characteristic of a
typical location in a
typical commercial or
hospital environment.