13
ECP Water Circulator Pump
If you have purchased the ECP with your System,
read the following to set up the ECP with the
FTIR600 stage. Refer to the ECP manual for
more details.
When heating the stage above 300°C for a pro-
longed period of time, the metal casing body of
the stage can get quite hot. Connect the tubing as
shown in the pictures below to cool the Stage.
FTIR600 Stage with water cooling tubing connec-
tions
FTIR600 Vertical Stage with water cooling tubing
connections.
Note:
If you have an LNP95 Liquid Nitrogen
Cooling system, the tubing from the Dewar must
NOT be connected to the FTIR600 stage when
heating above 300°C. The thin black capillary
tube will melt inside the heater and damage it.
Bypass cooling
ECP Inlet
ECP Outlet
ECP Inlet
ECP Outlet