Linkam Scientific Instruments CCR1000 Скачать руководство пользователя страница 12

 

12 

Load Sample Filter  

The  sample  filters  come  as  a  sheet  of  alumina  /
silica (Al

2

O

3

 96%/SiO2 4%) Non-Woven Fabric.   

A small disc of this fabric is needed to place the 
sample  on.    The  following  procedure  must  be 
used to load the filter into the Sample Holder. 
 

Warning

:  The  above  mater ial  is  har mful  by 

inhalation and when in contact with skin.  Possi-
ble  risk  of  irreversible  effects.    Wear  suitable 
protective  clothing.    Wear  suitable  gloves,  eye/
face protection or work in a fume cupboard.  This 
material is NON-CARCINOGENIC. 
 

Important  Note  Before  Heating  Sample: 

The 

Alumina/Silica Fabric filter on which the sample 
is  loaded  should  be  heated  up  to  600ºC  before 
doing  a  sample  run.   Either heat the  fabric in  an 
autoclave or using the CCR1000.  Remove the lid 
window and run gas through the stage during the 
heating phase.  This is so that any dust or debris 
in the stage or fabric filter is burned off and car-
ried away before start of experiment. 
 
Turn the Sample Holder upside down and with a 
forefinger  and  thumb  gently  rest  it  over  the  alu-
mina/silica fabric (1).  Once the sample holder is 
in position, apply a very light force and rotate the 
sample holder into the alumina/silica Non-Woven 
Fabric.  This will cut the fabric to fit the sample 
holder.   
 
Gently lift the sample holder away from the fab-
ric.  The filter will now be the perfect size for the 
sample holder. 
 
The  filter  is  not  yet  low  enough  in  the  sample 
holder and needs to be pressed into position.  This 
can  be  achieved  by  the  using  the  Filter  Load/
Removal Tool (2). 
 
The  filter  should  be  in  the  sample  holder  and 
ready to be pressed into position as shown above.  
Make sure that the tool is round the correct way to 
press the filter into the sample holder.   

 

Note: To remove the filter, the tool can be used to 
push it out of the sample holder.  Balance the tool 
on  a  flat  surface  so  that  the  smaller  diameter  of 
the tool is protruding upwards.  Hold the sample 
holder  over  the  tool  (the  Filter  Load/Removal 
Tool can be supported by hand if required).  Gen-
tly  lower  the  sample  holder  over  the  filter  load/
removal tool. 

Содержание CCR1000

Страница 1: ...CCR1000 Catalyst Cell Reactor Linkam Scientific Instruments USER GUIDE...

Страница 2: ...cope with Dovetail Substage 7 Set Up the CCR1000 8 Electrical Cable Connectors 8 Connection to T95 System Controller 9 Connect the ECP Water Circulator Pump 10 Connect the Gas Lines 10 Pressure Warnin...

Страница 3: ...with us to Activate your warranty and technical support Access the online setup videos Permanently unlock the Linksys32 software if purchased If you have purchased Linksys32 software please install t...

Страница 4: ...isuse operation outside of the environmental specifications for the product or improper site preparation or maintenance Technical Support Any technical questions or queries should be addressed to the...

Страница 5: ...oors 9 Each product is equipped with a 3 wire grounded earth mains plug or a free end 3 wire mains lead The plug only fits into a grounded type outlet The free end mains lead should be connected to a...

Страница 6: ...y porous ceramic cloth filter Gas is introduced via a high pressure valve and is then heated by sample holder before passing through the sample A high pressure outlet valve can be connected to various...

Страница 7: ...base of the stage using the supplied hex screws and the outer most holes in the base plate Adjust the two positioning screws 1 and the thumb screw 2 so that approximately 5mm of thread is exposed on...

Страница 8: ...d with the CCR1000 1 CCR1000 2 Green Temperature Sensor Cable 3 Power Cable 4 Stage Connection Lead Electrical Cable Connectors Plug in the green Temperature Sensor cable to stage Make sure the and ma...

Страница 9: ...95 System Controller From the stage connect the green Temperature Sensor cable to the Stage Socket 1 and connect the Stage Connection Lead to Stage Connection Socket 2 Refer to the T95 LinkPad Manual...

Страница 10: ...ocate the gas line and connector into the stage body No force is necessary Now screw in the connector until it stops using only fingers No spanner is necessary unless using stainless steel tubing Note...

Страница 11: ...Holder 2 The next step requires great care and minimal force Once the tool is axially aligned with the sample holder it will need to be pressed over the outer diameter of the Sample Holder Rotate the...

Страница 12: ...ment Turn the Sample Holder upside down and with a forefinger and thumb gently rest it over the alu mina silica fabric 1 Once the sample holder is in position apply a very light force and rotate the s...

Страница 13: ...to the central hole in the sample loading base until resistance is felt Excessive forces are not needed as the sample holder should slide into the sample loading base Place the Sample Loading Funnel 3...

Страница 14: ...Holder into the heater assembly Rotate the sample holder to make sure that it is axially aligned with the heater assembly Once the sample holder is positioned in the heater assembly place a fore finge...

Страница 15: ...nt Spares Kit CCR SLK Sample Loading Kit CCR CC Ceramic Cloth for Sample Loading 6x3 5cm x4 CCR CPF Cole Palmer Fitting with Ferrule x4 W22Q1 22mm Diameter Quartz Window 1mm thick x2 CCR ALSH Alumina...

Страница 16: ...am Scientific Instruments Ltd Tel 44 0 1737 363 476 Fax 44 0 1737 363 480 Email support linkam co uk Unit 8 Epsom Downs Metro Centre Waterfield Tadworth Surrey KT20 5LR UK www linkam co uk Version 1 0...

Отзывы: