LINET SELLA Скачать руководство пользователя страница 44

 

D9U004SLA-0101 

44/49 

 

We recommend using cleaning and disinfection aids based on: 

 

  quaternary ammonia compounds 
  aldehyde preparations 
  peroxide compounds 
  preparations with active chlorine 

 

 

We  recommend  using  cleaning  and  disinfection  aids  from  renowned 
producers such as: 
 

  Bochemie 
  B.Braun Medical 
  Bode Chemie  
  ECOLAB  
  Johnson Diversey 

 

Hygiene  products  have  been  tested  at  the  highest  recommended 
concentrations. 

 

It  is  always  necessary  to  follow  the  manufacturer’s  instructions  for  individual 
cleaning and disinfecting agents.  
 
BORCAD  Medical  a.s.  is  not  responsible  for  damage  to  cushioning  (e.g. 
cracking,  paint  flaking,  etc.),  which  would  result  from  improper  use  of 
disinfectants. The warranty is not valid in such cases.  
 
It  is  crucial  to  cover  the  cushioning  before  every  use  with  a  clean  blanket  for 
hygienic reasons.  

 

9 Maintenance 

  Treads  of  the  fastening  screws  or  the  joints  on  the  frame  parts  may  be 

degreased by disinfectants. In such cases, timely lubrication prevents costly 
repairs.  

  The  product  destined  for  disposal  or  repair  must  be  properly  cleaned  and 

disinfected.  

 

Содержание SELLA

Страница 1: ...Multifunctional Chair SELLA UN PACKING MANUAL INSTRUCTION MANUAL D9U004SLA 0101 Version 02 Publication Date 2022 07...

Страница 2: ...RCAD Medical a s Related links http www linet com cs borcad medical All rights reserved All trademarks and brands are the property of the appropriate owners The manufacturer reserves the right to chan...

Страница 3: ...33 7 Accessories 34 7 1 SLA ZP 037 Armrests standard pair 34 7 2 SLA ZP 038 Armrests for blood donation pair 34 7 3 SLA ZP 001 Table 35 7 4 SLA ZP 024 Table with safety barrier 35 7 5 SLA ZP 023 Table...

Страница 4: ...004SLA 0101 4 49 9 Maintenance 44 9 1 Technical specification 45 9 2 Transportation storage and operation conditions 46 9 4 Service repairs 47 9 5 Environmental protection and disposal of the product...

Страница 5: ...always only in accordance with these instructions Please keep these instructions always near the chair throughout its operation course Every person operating the chair must read and understand the con...

Страница 6: ...operation i e if the product has been continuously operated for time period x then it must rest idle for time period y For example int 10 20 means that after 10 minutes of continuous use positioning t...

Страница 7: ...sportation marking Relative air humidity Transportation marking Air pressure Marking according to the directive EC 2002 96 EC Directive about the disposal of old electric and electronic appliances Sym...

Страница 8: ...A Power intake units VGA Computer standard for computer imaging technology USB Universal serial bus LCD Liquid crystal display IT Information technology PC Personal computer dB Sound intensity unit hP...

Страница 9: ...facturing Date Year Month Day 4 DI Device Identifier GTIN Global Trade Item Number 5 PI Product Identifier 6 2D Bar Code GS1 DataMatrix DI PI UDI 7 1D Bar code GS1 128 Serial number 8 Symbols Any pote...

Страница 10: ...D9U004SLA 0101 10 49 1 Un packing manual Cut the tapes and remove the upper cover Remove all boxes with accessories and the accessory equipment from the bottom lid...

Страница 11: ...11 49 D9U004SLA 0101 Remove the equipment if it is a part of your order Telescopic infusion stand Side support pair...

Страница 12: ...9 Pull out the chair from the bottom slots and move out of the bottom lid Unless the chair is completely assembled you may not use the lever for adjusting the backrest or the lever for adjusting the b...

Страница 13: ...9 D9U004SLA 0101 Installation of gas spring for SLA IB version Put gas spring on the back section and remove protective plastic tube Put a rope on the control head it helps when you push the control h...

Страница 14: ...tective caps from the gas spring and draw bar Use the pins to connect the gas spring lug to the fork at the bottom part of the backrest Define dimensional clearances using plastic washers and secure t...

Страница 15: ...e backrest lug Similarly connect the draw bar lug with the backrest fork situated on the other side of the chair 1x PE washer thickness 1 mm 1x PE washer thickness 2 mm Gas spring lug 1x PE washer thi...

Страница 16: ...49 Undo the leatherine connection at the fork where the gas spring lug is fixed and move the cable into correct position Cut the tape securing the gas spring head under the seat Correct cable position...

Страница 17: ...he seat into the chair frame Pay attention to the seat orientation Connect the infusion rod holder to the chair if it is included First remove the infusion holder plug Then screw the infusion holder t...

Страница 18: ...ing off 6 Plastic chassis cover 7 Front wheels 7a small 7b large 8 Rear wheels 9 Headrest adjustable for height 10 Handle of the back part with a hook Control lever for adjusting bearing area 11 Multi...

Страница 19: ...X SLA CX SLA DX Fixed chassis with large front wheels Fixed chassis with front turning wheels Chassis with lift and turning wheels SLA IX SLA JX Chassis with lift and front large wheels and getting of...

Страница 20: ...D9U004SLA 0101 20 49 Color versions of cushioned parts X Type Color X A Apricot orange B Baltic blue C Cr me F Schwarz black D3 Pastel orange...

Страница 21: ...abel containing the name and brand of the manufacturer serial number or batch ID and date of manufacture Any serious incident that has occurred in relation to the device should be reported to the manu...

Страница 22: ...Note Danger of frostbite by contact of CO2 with human skin Do not use halotron extinguisher in closed rooms without ventilation danger to your health Use Do not use Powder extinguisher Foam extinguish...

Страница 23: ...ndependence 4 1 User population Suitable in a variety of clinical settings including nursing homes ICU emergency one day care acute care med surge geriatrics and rehabilitation centers by adult patien...

Страница 24: ...lowering to the lowest position no objects that could collide with the chair or block it may be present nearby When stopping the chair it must be braked using the central brake alternatively the rear...

Страница 25: ...25 49 D9U004SLA 0101 5 1 Basic positions of the chair...

Страница 26: ...ed with four wheels The rear small wheels rotate around their axis and are braked simultaneously using the central brake versions SLA AX SLA CX SLA DX SLA IX a SLA JX Braking is initiated by stepping...

Страница 27: ...they have a brake but allow movement of the chair on uneven surface balconies terraces transfer between buildings etc Small front wheels turn around the axle and have no brake it is possible to set f...

Страница 28: ...h lift SLA DX IX JX When the chair is transformed into the bed shape or position it may be adjusted to the necessary height by repeated stepping on the lifting pedal which is situated on both sides of...

Страница 29: ...sing the incorporated gas spring When the chair is transformed into the bed shape or position it may be altered easily using the incorporated gas spring By pressing or pulling the handle optimal posit...

Страница 30: ...arm support and can be adjusted to four positions They are cushioned on the upper part Positions are altered by pressing the armrest lever and simultaneous adjusting to the desired position Transport...

Страница 31: ...support for the feet during sitting and transporting during getting on off it acts as an active security element and the projecting side edges protect the feet during transportation and eliminate risk...

Страница 32: ...ay lever as far as the click in the direction of the arrow Basic position Activation of getting on off help By repeated stepping on the lifting pedal the bed first inclines by 25 forward and then it s...

Страница 33: ...33 49 D9U004SLA 0101 6 8 Hook An integral part of the driving handle is a shifting hook of bearing capacity up to 20 kg...

Страница 34: ...pulling up the arm rest may be set to higher position Horizontal adjustment Each arm rest may be adjusted horizontally to the side by simple lifting of the arm rest and rotating it to the side There...

Страница 35: ...side when used as a side support is 40 Kg The table can t be used with SLA ZP 038 Armrests for blood donation 7 4 SLA ZP 024 Table with safety barrier The maximum loading of the table from above and...

Страница 36: ...SLA 0101 36 49 7 5 SLA ZP 023 Table safety lock The pins of the table are inserted in the front openings in the multifunctional armrests It is possible to adjust two secure positions more and less ins...

Страница 37: ...Procedure 1 Because of safe installation always slip the table into the holder with stainless steel interposer bars directing to the right Fig 1 2 First slip the upper bar into the upper holder by mov...

Страница 38: ...emoving the plastic plug the stand is screwed in the left or right opening over the rear wheel 7 9 SLA ZP 003 Telescopic Infusion stand The infusion stand is inserted in the holder SLA ZP 002 it is im...

Страница 39: ...39 49 D9U004SLA 0101 7 10 SLA ZP 011 10 X Side support pair 7 11 SLA ZP 010 X Headrest big 7 12 SLA ZP 013 X Prolonged stripe for headrest...

Страница 40: ...D9U004SLA 0101 40 49 7 13 SLA ZP 019 Urine bag holder 7 14 SLA ZP 119 Urine bag holder If you use this type of Urine bag holder then you can t use side guards at the same time...

Страница 41: ...y adjust the position of the back part After releasing the lever the position is fixed 7 16 SLA ZP 006 Independent adjustment of legrest The position of leg rest only may be altered by pressing the co...

Страница 42: ...on only Be especially careful when transporting the chair and the oxygen bottle The oxygen bottle may fall out when the back side is reclined and upon incautious manipulation 7 18 SLA ZP 044 X Patient...

Страница 43: ...mmend using standard preparations for desinfection of surfaces based on for example quaternary ammonia compounds aldehyde preparations peroxy compounds and preparations containing active chlorine The...

Страница 44: ...o follow the manufacturer s instructions for individual cleaning and disinfecting agents BORCAD Medical a s is not responsible for damage to cushioning e g cracking paint flaking etc which would resul...

Страница 45: ...mm Positioning angle Backrest positioning angle 12 70 12 70 12 70 Seat positioning angle 10 10 10 15 Trendelenburg position Leg support positioning angle 11 87 11 87 11 87 Armrest positioning angle 2...

Страница 46: ...conditions Packed chairs may be exposed for a period of up to 15 weeks in conditions that do not exceed the following limits Transport and storage conditions 10 C 40 C 30 75 795 hPa 1060 hPa Operatio...

Страница 47: ...vice 673 739 45 Fry ovice Czech Republic Tel 420 558 640 632 420 736 628 209 420 736 628 211 Fax 420 558 668 087 e mail service borcadmedical com www borcadmedical com For repairs only use original BO...

Страница 48: ...respective directives Dispose of the packaging materi l according to the symbols and by delivering it to an authorised person The product consists of recyclable steel plastic and electronic component...

Страница 49: ...49 49 D9U004SLA 0101...

Отзывы: