13
3. Turn on the electrical equipment to measure the moment start up current. The
meter is locked to the highest measurement range.
4. Press select button for 1 second to exit surge current mode.
Resistance measurement
(See fig 1)
1. Insert red test lead into the V
Ω
Hz terminal and black test lead into the COM
terminal.
2. Set the rotary switch to
Ω
position.
3. Push select button to select resistance function. Displays shows
Ω
.
4. Connect the test leads across with the object being measured. The measured
value shows on the display.
Note
• The test leads can add 0.1
Ω
to 0.3
Ω
of error to resistance measurement. To
obtain precision readings in low-resistance measurement, that is the range of
200
Ω
, short-circuit the input terminals beforehand and record the reading
obtained. This is the additional resistance from the test lead.
• OL displays when the circuit is open or the resistor value is higher than max
range.
Frequency and Duty Cycle measurement
(See fig 2)
1. Insert red test lead into the V
Ω
Hz terminal and black test lead into the COM
terminal.
English