Suzhou Lily Tech Co., Ltd.
Page 4, Total 8 Pages
Parameter Table
Code
Function
Range
Unit
Indication
Default
T10 Temperature control working mode
0 ~ 2
0: R1 up/down limit control
1: External SSR, up/down limit control
2: External SSR, PID control
0
T11 Temperature set point
0.0 ~ 100.0
℃
/
℉
37.8
T12 Temperature hysteresis for R1 (T10 = 0)
0.0 ~ 20.0
℃
/
℉
Relative to T11, absolute value: T11 - T12
0.2
T13 Temperature hysteresis for SSR (T10 = 1)
0.0 ~ 20.0
℃
/
℉
Relative to T11, absolute value: T11 - T13
0.3
T14 Min stop time for temperature load (R1)
0 ~ 999
sec
1
T20 Temperature calibration
-9.9 ~ +9.9
℃
/
℉
0.0
T21 High temp. alarm relative point
0.
0 ~ 65.0/dis
℃
/
℉
dis: disable warning.
Absolute point = T11 + T21
dis
T22 Low temp. alarm relative point
0.
0 ~ 65.0/dis
℃
/
℉
dis: disable warning.
Absolute point = T11 - T22
dis
T25 High temp. protection relative point
0.
0 ~ 65.0/dis
℃
/
℉
dis: disable warning.
Absolute point = T11 + T25
Ref.: U20 = 1/2/3, U30 =
1/2/3
, U40 =
1/2/3
0.2
T26 Hysteresis for T25
0.0 ~ 65.0
℃
/
℉
Absolute point = T11+T25
-T26
Ref.: U20 = 1/2/3, U30 =
1/2/3
, U40 =
1/2/3
0.1
T30 Proportion of SSR PID
0.1 ~ 5000
Power/
℃
Kp for PID
39.3
T31 Integration of SSR PID
0 ~ 5000
sec Ti for PID. If Ti = 0, shut integration.
15.7
T32 Differential of SSR PID
0 ~ 5000
sec Td for PID. If Td =0, shut differential.
40.5
H10 Dehumidify / humidify mode
0 / 1
0: dehumidify, 1: humidify
1
H20 Humidity set point
0 ~ 99
%RH
50
H21 Humidity hysteresis
0 ~ 50
%RH Relative to H20, absolute value: H20 - H21
5
H22 Humidity calibration
-9.9 ~ +9.9
%RH
0.0
H23 Minimum stop time for humidity load (R3)
0 ~ 30
min
0
H24 High humidity warning relative point
0
~ 80/dis
%RH
dis: disable warning.
Absolute point = H20 + H24
dis
H25 Low humidity warning relative point
0
~ 80/dis
%RH
dis: disable warning.
Absolute point = H20 - H25
dis
H26 High humidity protection relative point
0 ~ 80/dis
%RH
dis: disable warning.
Absolute point = H20 + H26
Ref.: U20 = 1/2/3, U30 =
1/2/3
, U40 =
1/2/3
5
H27 Hysteresis for H26
0
~ 80
%RH
Absolute point = H20-H26
-H27
Ref.: U20 = 1/2/3, U30 =
1/2/3
, U40 =
1/2/3
2
U10 Left turn (R6) time unit
1 ~ 2
1: minute, 2: hour
1
U11 Left turn (R6) time of the timer
1 ~ 999
120
U12 Right turn (R7) time unit
1 ~ 2
1: minute, 2: hour
1
U13 Right turn (R7) time of the timer
1 ~ 999
120
U14 Turning times
0 ~ 999
0: unlimited turning times
0
U16 Turning motor working mode
0 / 1
0: Either R6 or R7 will be energized.
Need position switches, position arrival shuts off
the power supply to motor
1: Only energized during turning moment (U17).
Manual egg turning support
1
U17 Motor being energized time for every turning
0 ~ 999
sec Effective only when U16 = 1
30
U18 Hatching days
0 ~ 99
day 0: disabled hatching days counting
0
U20 Output R5 (air exhaustion) control mode
0 ~ 3
0: Air exhaust
1: Air e humi./temp.
high
limit protection
2: Air e temperature
high
limit protection
3: Air e humidity
high
limit protection
1
U21 Time unit for air exhaust period (R5)
0 ~ 2
0: second, 1: minute, 2: hour
1
U22 Air exhaust period (R5)
1 ~ 999
30
U23 Time unit for air exhausting time (R5)
0 ~ 2
0: second, 1: minute, 2: hour
0
U24 Air exhausting time (R5)
1 ~ 999
30
U30 Output R4 (Lamp or protection) control mode
0 ~
3
0: illumination control
1
: temperature and humidity
high
limit protection
2
: Temperature
high
limit protection
3
: humidity
high
limit protection
0
U34 Output R4 delay time for automatic shutdown
1 ~ 999
Min Only effective when U30 = 0
2
U40 Output R8 (alarm or protection) control mode
0 ~
3
0 = alarm output
1 = temperature and humidity
High
limit protection
2 = temperature
high
limit protection
3 = humidity
high
limit protection
0
U50
Door switch mode
0 / 1
0: Normal close, 1: Normal open
0
U65
Buzzing warning option
0 / 1
0: disable warning, 1: enable warning
1
U90 Password
000 ~ 999
000 = no password
0
U93 PID self auto tune power rate
10 ~ 100
%
Percent of heater’s full power rate
50