4.Advanced User Guidebook
25
DC component or offset can cause incorrect magnitude values of FFT waveform. To
minimize the DC component, choose AC Coupling on the source signal.
To reduce random noise and aliased components in repetitive or single-shot events, set
the oscilloscope acquisition mode to average.
What is Nyquist frequency?
The Nyquist frequency is the highest frequency that any real-time digitizing oscilloscope
can acquire without aliasing. This frequency is half of the sample rate. Frequencies above
the Nyquist frequency will be under sampled, which causes aliasing. So pay more
attention to the relation between the frequency being sampled and measured.
Use Vertical Position and Scale Knobs
1.
The
Vertical Position
knob is used to adjust the vertical positions of the
waveforms.
The analytic resolution of this control knob changes with the vertical division.
2.
The
Vertical Scale
knob is used to regulate the vertical resolution of the wave
forms. The sensitivity of the vertical division steps as 1-2-5.
The vertical position and vertical resolution is displayed at the left bottom corner of
the screen (see
Figure 4-1 Information about Vertical Position
How to Set the Horizontal System
The
HORIZONTAL CONTROLS
includes the
Horizontal HOR
button and such
knobs as
Horizontal Position
and
Horizontal Scale
.
1.
Horizontal Position
knob: this knob is used to adjust the horizontal positions of