Before lifting, keep the following points in mind:
• A responsible person within your organization should decide on a case-by-case basis whether one or more
caregivers are needed.
• Ensure that the sling selected meets the patient’s needs with reference to model, size, fabric and design.
• Before lifting, plan the lifting operation so that it can be done as safely and smoothly as possible.
• Before lifting, check that the lifting accessory hangs vertically and can move freely.
• Although Liko’s sling bars are equipped with safety latches, special caution must be exercised: Before the patient
is lifted from the underlying surface, but after the straps have been fully extended, make sure the straps are
properly connected to the sling bar.
• Make sure the patient is sitting securely in the sling before transferring to another location.
• Never lift a patient higher off the underlying surface than is needed to complete the lifting and transfer procedure.
• Never leave a patient unattended during a lifting situation.
• Make sure that the wheels on the wheelchair, bed, gurney, etc., are locked during the lifting/transfer operation.
• Always work ergonomically.
Safety Instructions
Liko’s products undergo continuous development, which is why we reserve the right to make product changes
without prior notice. Contact your Liko representative for advice and information about product upgrades.
Care and Maintenance
Check the sling regularly, especially after washing. Check the following points with regard to wear and damage:
• Fabric
• Straps
• Seams
• Suspension loops
Do not use damaged lifting accessories.
If anything is unclear, please contact the manufacturer or supplier.
Medical Device Class I
Liko Soft Original HighBack Sling has been tested by an accredited testing institute and complies with
all requirements for MDD Class I products and the harmonized EN ISO 10535 standard.
IMPORTANT!
Lifting and transferring a patient always involves a certain risk. Read the instruction guide for both the lift
and lifting accessories before use. It is important to completely understand the contents of the instruction
guide.The equipment should be used only by trained personnel. Ensure that the lifting accessories are suitable
for the lift being used. Exercise care and caution during use. As a caregiver, you are always responsible
for the patient’s safety. You must be aware of the patient’s ability to handle the lifting situation.
Washing instructions:
See the sling’s product label.
Liko is quality certified in accordance with ISO 9001 and its equivalent for the medical device industry,
ISO 13485. Liko is also certified in accordance with environmental standard ISO 14001.
Design and Quality by Liko in Sweden
Expected Life Time
The product has an expected life time of 1-5 years or more during normal use. Expected life time varies depending
on, for instance, fabric, use frequency, washing procedure, and weight carried.
Periodic Inspection
The product must be inspected at least once every 6 months. More frequent inspections might be required if the
product is used or washed more often than normal. See Liko’s protocol Periodic Inspection for the respective
product.