background image

© Cop

yrigh

t Lik

o R&D AB 2017

Distributor:

Liko AB 

Nedre vägen 100

SE-975 92 Luleå, Sweden

Manufacturer:

Balea

8, avenue du Grand Chene

F-34270 Saint-Mathieu-de-Tréviers, France

www.hill-rom.com  I  www.liko.com

Maximum Load

The Scal’Up

®

 scale must have a maximum capacity less than or equal to the lift unit’s maximum capacity. 

Example:  

A lift unit that is approved for 200 kg must be used together with a scale that is approved for at most a maximum load of 200 kg.  

A lift unit that is approved for 205 kg must be used together with a scale that is approved for at most a maximum load of 200 kg. 
For the assembled lift unit, including accessories, the maximum load is always the lowest maximum load rating for any of the 

components. 
Study the markings on the lift and lifting accessories or contact your Hill-Rom representative if you have any questions.

Technical Data

Accuracy class:    

Class III according to the requirements of the NAWI 2014/31/EU Directive.

Readout: 

 

 

Sensitive to 0.1 kg

Tolerance:  

 

2-50 kg ± 0.05 kg 

     

 

50-200 kg ± 0.1 kg

     

 

200-300 kg ± 0.15 kg

Max. load:

  

160 kg/175 kg/200 kg/230 kg/250 kg/300 kg

Min. load: 

  

2 kg 

Reset: 

 

 

One button reset ( 

0/T

 )

Power supply: 

   

Battery 9 V 

Battery:

  

 

9 V Alkaline (PP3, MN1604, 6LR61, E-block). 

     

 

Battery life: 6 months

Display: 

 

 

Digital, liquid crystal (LCD)

Size:

  

 

173 x 60 x 100 mm (L x W x H)

Weight: 

  

0.9 kg

Temperature range: 

 

+10

C to +40

C

Humidity range: 

 

0 % - 80 %