background image

 www.hillrom.com

7EN161113 Rev. 5  •  2020

10

Disinfection Instructions

1. For the use of suitable disinfectants see chapter “Use of cleaning and disinfecting agents on LikoGuard™ Overhead lift“. 

2. Use the choice of disinfectant according to the manufacturer’s instructions and repeat the work step as in ”Cleaning instructions.

3. Remove traces of disinfectant after disinfection. Wipe off the entire sling bar with a cloth moistened with clean water starting  

  from the top and working down. The cloth shall not be so damp that it drips.

Cleaning and Disinfection

Equipment

• 

 Safety equipment (such as rubber gloves, protective glasses, aprons, face masks/protective masks and overshoes) according 

to guidelines and routines in your organization and the manufacturer’s instructions.

• 

Clean buckets

• 

Cloths

• 

Soft brushes

• 

Warm water

• 

 See chapter ”Use of cleaning and disinfecting agents on LikoGuard™ Overhead lift” for choice of suitable cleaning and 

disinfecting agents.

All parts must be kept clean in order to avoid the risk of infection. Necessary precautions must be taken to eliminate all visible 

stains and dirt. The cleaning and disinfection methods described are specifically aimed at the lift and all accessories. It has 

been designed to save time and contribute to effectively fighting healthcare related infections.

SlingGuard™450 sling bar

Lift Strap

Latch

Cleaning instructions

 

1. If the sling bar includes the Quick-hook Multi, remove the Sling bar by disconnecting the Quick-hook Multi from the lifts      

    

Multi-Link connector, see ”Assembly of SlingGuard™ Sling bar with Quick-Hook Multi”.

2.  Clean the sling bar with a cloth moistened with warm water and a neutral cleaning agent approved by your organization. 

 

A soft brush may be used to remove stains and resistant dirt.

3.   Wipe off the entire sling bar with a cloth moistened with clean water starting from the top and working down. The cloth 

shall not be so damp that it drips.

IMPORTANT! 

The lift strap or the Emergency stop cord shall not be exposed to following agents: CSI, Oxivir Tb, Dispatch, 

Chlor-Clean, Dismozon Pur or the equivalent.

Cleaning and disinfection must always be carried out in accordance with the recommended instructions.

Содержание 3308020

Страница 1: ...450 Prod No 3308020 SlingGuard 450 with Quick Hook Multi Prod No 3308520 SlingGuard 670 Twin Prod No 3308040 SlingGuard 670 Twin with Quick Hook Multi Prod No 3308540 7EN161113 Rev 5 ENGLISH 2020 Sli...

Страница 2: ...ed assembly of SlingGuard sling bar to lift 8 Assembly of SlingGuard sling bar with Quick Hook Multi 9 Dimensions 9 SlingGuard 450 sling bars 9 SlingGuard 670 Twin sling bar 9 Cleaning and disinfectio...

Страница 3: ...tructions for use before use Date of manufacturing Important Failure to follow the instructions may lead to damage to the product Caution consult instructions for use Read instructions for use before...

Страница 4: ...the lift Fixed attachment or the quick attachment device Quick Hook Multi SlingGuard Sling bars has latches to protect against involuntary unhooking of the slings suspension loops Medical Device Clas...

Страница 5: ...106 kPa In the event of any queries regarding this product contact your Hill Rom representative For guidance and support in choosing the correct sling visit www hillrom com or contact your Hill Rom r...

Страница 6: ...e for the selected sling before lifting Attach lifting accessories correctly so that the sling bar hangs level and all sling bar hooks take all the weight when lifting Before lifting always ensure tha...

Страница 7: ...g bar hooks Note For correct and safe use of a sling always study the instruction for use for the selected sling before lifting Sling with 4 suspension loops Upper strap loops Leg support loops A A B...

Страница 8: ...t Fits LikoGuard L Overhead Lift Prod No 3301030 LikoGuard XL Overhead Lift Prod No 3301040 Torque 1 25 0 25 Nm 11 2 2 in Ibf 1 2 3 5 4 6 7 Tools required Torque Wrench 0 10 Nm 88 in Ibf Torx T25 X 2...

Страница 9: ...b Installation dimensions the distance between the sling bar hooks attachment point for sling etc to the sling bar s central suspension point CSP to lift A B C H 200 757 133 171 A B C H 7 9 29 8 5 2...

Страница 10: ...o avoid the risk of infection Necessary precautions must be taken to eliminate all visible stains and dirt The cleaning and disinfection methods described are specifically aimed at the lift and all ac...

Страница 11: ...Alcohol Isopropyl alcohol 70 5 0 7 0 Viraguard or equivalent Veridien Quaternary ammonium n Alkyl dimethyl benzyl ammonium chlorides 0 105 n Alkyl dimethyl ethylbenzyl ammonium chlorides 0 105 11 5 12...

Страница 12: ...and any required cleaning and disinfection of the medical device after use and prior to its disposal Customers should adhere to all federal state regional and or local laws and regulations as it perta...

Отзывы: