LD14x
absolute (decimal point lit solidly) to incremental
(blinking decimal point) measurement and vice
versa.
Zero setting in incremental modes (see 6.2.1 Set
datum (reference)) does not change the absolute
value in the background.
The function is enabled only if
set to
"yES"
.
6.2.3 Mm/inch display modes
Mm/inch display mode can be changed by pressing
key for 3 sec. The function is enabled only if
parameter is set to
"yES"
.
6.2.4 Changing the Offset value
Press
P
and
keys simultaneously to display 1.
Offset value (
). Use
and
keys to change
the value and save with
*
Offset values can be changed only in setup menu.
Offset function is enable if
to
“yES”
.
and
values.
6.2.4.1 Fractional offset display
The fractional inch display mode allows to set offset
values (
OFS
) in the following way:
- 1
st
digit blinking
increases
1
/
64
” pressing
key.
- 2
nd
digit blinking
increases
1
/
32
” pressing
key.
- 3
rd
digit blinking
increases
1
/
16
” pressing
key.
- 4
th
digit blinking
increases
1
/
8
” pressing
key.
- 5
th
digit blinking
increases 1” pressing
key.
- 6
th
digit blinking
increases 10” pressing
key.
6.2.5 Datum modification
Press
P
and
keys simultaneously to display datum
. Use
and
keys to change the value
and save with
*
key.
This function is enabled only if
set to
“yES”
.
6.3 Setup / Parameter setting
Press
P
key for 3 sec. to enter setup,
“SEtUP”
is
displayed.
Press
key to enter
MENU 1
(parameters)
Press
*
key to enter
MENU 2
(RS-232 serial
interface)
Press
P
key to access the next Parameter and
Parameter setting.
Press
P
key for 3 s to exit the setup at any point.
6.3.1 Default parameters (factory settings)
All default values are written in
BOLD
characters.
The display can be reset to default parameters with
the following procedure:
•
remove the battery and wait 10 sec.;
•
press
*
key (
“dEFPar”
is displayed) while
inserting the battery.
6.3.2 Parameter list MENU 1
Unit
Measurement unit [dEC, FrEE, dG1, dG2, IdEC, Ifrct]
It sets the measurement unit and the display mode.
dEC
= linear measurement display (decimal)
FrEE = display with conversion factor
dG1 = angular display (-
..-0.1°..0.0°..+0.1°..+
)
dG2 = angular display (..359.9°..0.0°..359.9°..0.0°..)
IdEC = inch display mode
Ifrct = fractional inch mode (eg. 12.31.64 = 12"
31
/
64
)
*
= save,
P
= next parameter,
P
for 3 s. = exit
COn
only with
= FrEE, dG1, dG2
It allows to set a free conversion factor to display
non-metric units or angles.
Value range:
FrEE = 0.00001 –
1.00000
dG1, dG2 = 0.00001 – 9.99999
Example 1
We want to display a 90° angle (from 0° to 90°) with
0.1° resolution on a round table having 785.4 mm
circumference.
The measurement length on 360° is 785.4 mm, thus
it is 785.4 / 4 = 196.35 on 90.0°.
= 900 : 19635 = 0.045836
Example 2
We want to display angles on a magnetic ring
having a 114.5 mm diameter.
The circumference is 114.5 * 3.14 = 359.53 mm
= 3600 : 35953 = 0.10013
*
= save,
P
= next parameter,
P
for 3 s. = exit
____________________________________________________________________________________________
MAN LD14x I_E_D 4.4.odt
11
www.lika.it