LightOx PhotoReact 365 Скачать руководство пользователя страница 14

New experiment

Rerun experiment

Settings

User Manual 

5.3. Performing an 

irradiation experiment

Running Experiment. Step 1.

Next, on the display start-up screen of the Control Unit, 
select either 

New experiment

 or 

Rerun experiment

After selecting 

New experiment

 enter the experiment 

name using the touch keyboard on the next screen. 

Select 

Save

 to bring up the experiment settings screen 

shown below. 

14

Содержание PhotoReact 365

Страница 1: ...Applies to PhotoReact 365 Model lightox co uk LightOx Limited 2020 All rights reserved User Manual...

Страница 2: ...7 18 20 21 22 22 23 23 23 24 24 25 26 No part of this publication may be reproduced transmitted transcribed stored in a retrieval system or translated into any language in any form by any means withou...

Страница 3: ...Data export as csv files Device has locking lid that reduces user exposure to harmful light rays Technical Specifications PhotoReact 365 Key Features The PhotoReact 365 is a benchtop instrument that...

Страница 4: ...ld only be used with the locking lid in place Do not bypass the lock activation switch as looking into the device when activated may result in damage to eyes or to exposed skin Only use the device wit...

Страница 5: ...n illumination using light sources that allow for controlled intensity and wavelengths to be delivered to the sample plate The PhotoReact 365 includes a UV sensor to monitor the health and light outpu...

Страница 6: ...required to take adequate measures If the PhotoReact 365 is used in a manner not specified within this manual the protection provided by the equipment might be impaired It is the responsibility of the...

Страница 7: ...The PhotoReact 365 is comprised of the following Base Unit Removable locking lid Control Unit USB cable connecting Control Unit to Base Unit Power Supply User Manual 4 System Components 7 2 2 3 3 1 1...

Страница 8: ...f space between the perimeter of the instrument and the nearest object The PhotoReact 365 is designed to work at ambient room temperature Transport the PhotoReact 365 in a horizontal position and avoi...

Страница 9: ...t Warning Risk of electrical shock Before plugging into mains power visually inspect the power supply plug and cable to ensure there is no damage If there are signs of damage do NOT use Note There are...

Страница 10: ...xperiment Rerun experiment and the system Settings button New experiment allows the user to input new settings Rerun experiment allows the user to reuse settings from a previously run experiment When...

Страница 11: ...the up and down toggle buttons to set the date day month and year and time hour minute and second in two digit format Select Save to save the user inputs into the memory Note once the date time is se...

Страница 12: ...ng the lid counter clockwise fully until it stops and lift the lid upwards with two hands Take the lid off place the multi well or tissue culture plate on top of the glass surface in preferred orienta...

Страница 13: ...Replace the lid ensuring the lid locking pins align with the round holes on the Base Unit Turn the lid clockwise fully to lock the lid into place Note The PhotoReact 365 will not function unless the l...

Страница 14: ...ing Experiment Step 1 Next on the display start up screen of the Control Unit select either New experiment or Rerun experiment After selecting New experiment enter the experiment name using the touch...

Страница 15: ...hen press the Hold button to the right of the top sliders Next use the middle slider to set the irradiance The total Energy delivered will be automatically calculated by the software Alternatively the...

Страница 16: ...UV sensors are displayed Note If the lid is removed during an experiment the LEDs will automatically turn off for safety but the fans will continue to run and the user will be prompted to re close th...

Страница 17: ...After selecting the name the next screen will allow the user to modify the parameters and experimental name by selecting the Modify button or to re run the experiment without changes by selecting Run...

Страница 18: ...5 5 Downloading experimental data Copy logs to flash drive Step 1 To download data files to a flash drive press the systems setting button on the start up screen On the settings screen press the Copy...

Страница 19: ...flash drive Step 2 Insert a USB drive into the Data Port on the back of the Control Unit next to the power connector Once inserted the Control Unit will automatically start downloading log files to th...

Страница 20: ...e all files are downloaded an Export complete prompt will display It s now safe to remove the USB drive and press the Done button to be taken back to the settings screen Note The file size for each lo...

Страница 21: ...nloading the files to the USB drive To clear all log files select the Clear all Logs button on the settings screen To permanently delete all log files from the controller press the Clear logs button C...

Страница 22: ...play Press the Back button to return to the settings screen Unplugging the unit or switching off the unit while it is running will cause no harm to the system nor will it cause an unsafe condition NOT...

Страница 23: ...onics with liquid cleaners If necessary remove all dust from external surfaces with dry compressed air In the event of a spillage follow local safety regulations PhotoReact 365 units should be dispose...

Страница 24: ...the power supply Q My UV sensor displays a reading over 100 during a test A It s possible the UV sensor will report a value slightly over 100 during the first few months of use The instruments are ca...

Страница 25: ...tions whichever occurs soonest Ambient conditions for operation Suitable for indoor commercial use only e g laboratories and testing stations Temperature 15 to 40 C 41 to 104 F Rel humidity 0 to 80 no...

Страница 26: ...LightOx Limited 2020 All rights reserved Contact LightOx Ltd Wynyard Park House Wynyard Avenue Wynyard Billingham England TS22 5TB tech support lightox co uk...

Отзывы: