1. Density Indicator
This display will show the density values. The density is formulated by intensity
times the duration in sec and divided by one thousand
2. Treatment Mode Selection Adjuster
This adjuster can be used to select preset treatment mode as Standard mode,
Intermediate mode, Accelerated Mode, and Custom mode. Its individual preset
configuration as follows:
a.
STANDARD:
3.0mW/ cm² x 30min exposure @ Total 5.4J/cm²
b.
INTERMIDIATE:
9.0mW/cm² x 10min exposure @ Total 5.4J/cm²
c.
ACCELERATED: 18mW/ cm² x 5min exposure @ Total 5.4J/cm²
d.
RAPID:
30mW/cm² x 3min exposure @ Total 5.4J/cm²
3. Intensity Indicator
This indicator will show the light intensity values. It ranges from 0 to 30mW/cm
2
4. Intensity Adjuster
The adjuster up / down key can be used to increase or decrease the UV light intensity.
5.
Duration Indicator
This indicator will show the duration set up values according to the user 's
preference. It is in second unit and ranges from 10 to 1800S.
6.
Duration Adjuster
The adjuster up / down key can be used to increase or decrease the duration times.
7.
Spot Sizes Indicator
This indicator will show the spot sizes in mm unit. It ranges from 4 to 11mm.
8.
Spot sizes Adjuster
The adjuster up / down key can be used to increase or decrease the spot sizes.