![Lightmed LIGHTLas SLT Deux-V Скачать руководство пользователя страница 4](http://html1.mh-extra.com/html/lightmed/lightlas-slt-deux-v/lightlas-slt-deux-v_quick-start-manual_1912189004.webp)
YAG & SLT Laser Settings
9
10
11
12
13
14
15
16
17
18
9
10
11
12a
12b
13
14
15
16
17
18
Magnification Changer
Turn knob to desired magnification.
Switch from YAG to SLT Mode
Offset (YAG mode only)
Turn dial to adjust the offset. Make sure the dial says + FOR
POSTERIOR OFFSET, or - FOR ANTERIOR OFFSET. This ranges
from -500 to +500.
12a - YAG Energy Setting
Turn to adjust for YAG energy setting.
12b - SLT Energy Setting
Turn to adjust for SLT energy setting.
Energy Reading
View your adjusted parameters here.
Test Button
Fires the laser while keeping the shutter closed. Safest way to
check laser output energy.
Set Button
Changes the laser from Standby to Treat mode. Push once
desired parameters are set.
Set Pulse
Change pulses from 1, 2, or 3 pulses per laser shot.
Total Counter
Top shows total number of shots. Bottom shows total energy
used. Record values after procedure.
Reset
Resets the counter.