ApexRp Operators Manual
47
248083480-1 Rev 3
Table 6-8 Alarm Threshold Registers
Setting the Alarm Threshold Value
The Alarm Threshold Value is set in the low register of the channels.
Each channel has independent threshold value registers. Since any or
all channels can be enabled for alarms at any given time, each
threshold value applies to the corresponding channel. Setting a value
for channel 1 as 100 will not affect channel 2 setting of, say, 500. See
Table 6-9.
Table 6-9 Alarm Threshold Registers set to 1000
Register Data Type
Description
45009
unsigned integer Threshold for Particle Channel 1 [high]
(smallest particle size starts here)
45010
unsigned integer Threshold for Particle Channel 1 [low]
45011
unsigned integer Threshold for Particle Channel 2 [high]
45012
unsigned integer Threshold for Particle Channel 2 [low]
45013
unsigned integer Threshold for Particle Channel 3 [high]
45014
unsigned integer Threshold for Particle Channel 3 [low]
45015
unsigned integer Threshold for Particle Channel 4 [high]
45016
unsigned integer Threshold for Particle Channel 4 [low]
Registers
Particle
Channel
Threshold
Value
45009 – 45010
1
1000
45011 – 45012
2
1000
Содержание ApexR02p
Страница 1: ...Operators Manual ApexRemoteWithPump A I R B O R N E P A R T I C L E C O U N T E R W I T H P U M P...
Страница 2: ......
Страница 10: ...ApexRp Operators Manual 9 248083480 1 Rev 3 Intentionally Blank Page...
Страница 14: ...ApexRp Operators Manual 13 248083480 1 Rev 3 Intentionally Blank Page...
Страница 22: ...ApexRp Operators Manual 21 248083480 1 Rev 3 Intentionally Blank Page...
Страница 38: ...ApexRp Operators Manual 37 248083480 1 Rev 3 Intentionally Blank Page...
Страница 73: ...ApexRp Operators Manual 248083480 1 Rev 3 72 Intentionally Blank Page...