5. MAINTENANCE
18
2.
Intake filter -
Clean as needed (usually annually or until it turns black.)
If
necessary, replace the filter.
•
The intake filter
door
is
on the left side of the concentrator. The intake filter is
b
ehind the Transom Filter. Using a small screwdriver, open the access door.
Unscrew the
filter core. Remove the intake filter.
•
The maintenance frequency of the filter is based on actual time used and the
environment.
R
eplace the filtration core when it turns black.
Check
more
frequently if in a
n environment that collects dusty easily and
clean
when
visibly dirty.
5.3 Cleaning the Optional Humidifier Bottle
•
Change the water in the humidifier bottle every day.
•
To Clean: Wash the humidifier bottle weekly using household detergent. Then
rinse it completely under running water and dry.
•
To Disinfect: Disinfect the humidifier parts by immersing them in a disinfecting
solution, or a solution of 1-part vinegar diluted with 10 parts water. Then rinse
it completely under running water and dry.
•
To remove and disassemble the humidifier bottle:
(1)
Disconnect the tubing from the humidifier bottle and the device
(2)
Remove the humidifier bottle from the holder on the device and
remove the cover from the humidifier bottle to clean and disinfect.
5.4 Oxygen Nasal Cannula
Refer to
the nasal cannula
manufacturer’s instructions for cleaning.