background image

The hour will now start flashing. Press and release 

or 

to

change the hour. Note that “AM” or “PM” 
changes when you pass 12:00. 
When the correct hour is on the display, 
wait five seconds.

1

Firmly press and hold the lancing device against your forearm or palm for a few seconds.

Wait until the skin surface under the AST

Clear Cap changes color (as blood collects beneath

the skin). This tells you there is enough blood flow for a good sample. 

Then press the release

button while continuing to apply pressure. Keep holding the lancing device against your skin
until a round drop of blood forms under the cap.

When sampling blood from your forearm or palm, make sure the drop of blood is 

of at least one

microlitre

(

actual size) before you release pressure and remove the lancing device.

3

Choose your sample site.

Forearm

Choose a fleshy area of the forearm away
from bone, visible veins and hair.

Palm

Choose a fleshy area on the palm below the
thumb or pinky finger. Select a spot with no
visible veins and away from deep lines.

2

Replace the blue
cap with the 
AST

Clear Cap.

1

Getting a blood sample from your forearm or palm

The OneTouch

®

UltraMini

Meter allows you to sample blood from your forearm or palm as well as your fingertip. Before using your forearm or palm

for sampling, please speak to your healthcare professional and refer to the owner’s booklet for complete instructions.

Forearm

Palm

Forearm

Palm

Abbreviated Procedure Guide

With the meter off, press and
release 

to turn it on. Your most

recent test result will appear on
the display.

Press 

to move to the previous result stored

in the meter. Then, press 

or 

to move

forward or backward through all of your
results. When you’re finished, press and 

hold 

for two seconds until the meter turns off.

Reviewing past results

The minutes will now start flashing.

Press 

or 

to change the minutes, then

wait five seconds.

2

Setting the time and date

The meter stores a maximum of 50 blood glucose test results. You can review the results stored in the meter memory, starting with the most recent.

Before using your meter for the first time, you should check and update
these settings. Start with the meter off, then press and hold 

for five

seconds until the start-up test screen
appears. After the test screen, the pre-set
time and date will appear on the display 
for five seconds.

Note: If a setting does not need to be updated, simply wait five seconds.
The meter display will automatically advance to the next setting.

Check AM/PM. If not correct, continue to press 

or 

until

the hour and “AM” or “PM” are correct.

The year will now flash. Press 

or 

to change the year. When the correct year is
on the display, wait five seconds.

3

With the month flashing on the

display, press 

or 

to change the month,

then wait five seconds.

4

The day will now flash. Press 

or 

to change the day, then wait five seconds.

Your time and date settings will appear on
the display for five seconds. Your settings
will be saved and the meter will then turn off.

5

Need help?

Call the OneTouch

®

Customer Care Line 

1 800 663-5521

or

Visit us at www.OneTouch.ca

Trademarks owned by Johnson & Johnson and used under license.

Rev. 2007/03

©2007 LifeScan Canada Ltd., Burnaby, B.C., V5C 6C6

AW 088861A

33716 OTUMini APG 088861A E  7/3/07  2:56 PM  Page 2