The top edge of the Amplifier provides for several other connections, as shown below.
1
2
3
4
Figure 3: Connecting the R40 Amplifier (top face)
1. Nonin Xpod: the circular connector allows for the connection of a Nonin Xpod pulse oximeter for measuring
SpO2.
2. Electro-Cap: the 25-way D-type connector allows for the connection of a standard Electro-Cap.
3. Aux1 and Aux2: these two 3.5mm jack connectors allow for the connection of standard transducers like Body
Position, Respiration Belts etc.
4. Patient Event: this 3.5mm jack connector allows for the connection of a standard Patient Event Thumb Switch.
All these connections are Type BF isolated. The conductive part of connectors and transducers
should not contact other conductive parts including earth. Always ensure that the transducer fitted
is suitable for a connection of this type.
2.3
Starting the system
To start the system proceed as follows:
y
Plug the PC into the mains supply.
y
Switch on the PC and wait for Windows to load.
y
Ensure R40 Amplifier is connected via the USB lead.
y
Launch the Trackit Plus Software application. Refer to the Trackit Plus Software User Manual for more details.
y
These procedures also apply following a mains interruption.
2.4
Shutdown of the system
At the completion of a study proceed as follows to shut down the system:
y
Stop the recording by pressing the ‘Stop Rec’ button
y
Exit the Trackit program.
y
Shut down Windows.
y
Switch off the PC and disconnect the mains supply.