8. 

User Control Menu and Navigation

 

The User Control Menu is located at the top-left region of the touchscreen interface and displays three menu 
options: 

Main

Results

, and 

Settings

 

8.1 

Main Menu 
When the RDS-1500 is powered on, the Main Menu will be displayed.  The user may return to the 
Main Menu at any time by selecting the 

Main

 tab from the User Control Menu.  From the Main 

Menu, the user may run a test. 

8.2 

Results Screen 
The RDS-1500 stores the results of each test conducted.  This data can be viewed at any time by 
selecting the 

Results

 tab from the User Control Menu.  The Results Screen displays the test name, 

sample ID, date and time, and result for each test.  Results can be sorted by category (Name, 
Sample ID, Time, or Result) by selecting the desired sorting category with the stylus.  Individual 
test images can be viewed by selecting the desired test from the list of test results, then selecting 

View

 

8.3 

Settings Screen 
Selecting 

Settings

 from the User Control Menu will bring up the Settings screen from which the 

user may edit the user list, run diagnostics, edit date and time, change sound settings, reset the 
program, and access other advanced settings. 
8.3.1 

To customize settings, select the appropriate check boxes for the desired displays: 
8.3.1.1  When checked, 

Display Line Density Values

 will display an exact value of the 

line densities for both the control and test lines in the result of a test. 

8.3.1.2  When checked, 

Display Ratios (Test/Control line)

 will enable the unit to show 

the ratio of the intensity of the test line to the intensity of the control line. 

8.3.1.3  When checked, 

Display European Date Format

 will display the date in the 

“Day-Month-Year” format. 

8.3.1.4  When checked, 

Auto-print

 will enable automatic printing of the results for each 

test that is run. 
Note:  A printer must be connected to the unit to use this feature. 

8.3.2 

To add, edit, or delete user names, from the Settings Screen, select 

User List

, make 

desired changes, then select 

Back

8.3.3 

If it is necessary to calibrate the stylus, from the Settings Screen, select 

Calibrate Stylus

.  

This will bring up the Stylus Properties screen. 
8.3.3.1  The double-tap speed can be set and tested by following the instructions under 

the 

Double-Tap

 tab. 

8.3.3.2  The 

Calibration

 tab allows the user to recalibrate the screen to properly respond 

to the stylus.  Selecting 

Recalibrate

 will bring up a screen that asks the user to 

carefully press and briefly hold the stylus on the center of the target, repeating as 
necessary to ensure calibration. 

 

9. 

Notification Icons

 

To the right of the User Control Menu, in the top-right corner of the display interface, are the three notification 
icons: 

9.1 

The Connection Status icon 

 is displayed when the RDS-1500 is connected to a computer via 

the USB cable.  This icon indicates that the unit is ready to transfer data to the desktop. 

9.2 

Power Properties icons display information regarding battery charge percentage and power 
scheme. 

9.2.1 

The AC power icon 

 is displayed when the unit is charging, connected to a 

computer, and is using the main battery. 

9.2.2 

The Battery icon 

 is displayed when the device is not charging and is using the 

main battery. 

Содержание AbraScan 475025

Страница 1: ...ation the reader and test strips provide semi quantitative and or qualitative evaluation of Microcystins in contaminated water samples Positive samples should be confirmed by HPLC GCMS or other conventional methods 2 Reader Features The AbraScan RDS 1500 Test Strip Reader has a color touchscreen interface and a rugged IP67 compliant design and features a rechargeable internal lithium ion battery e...

Страница 2: ...y life and ensure optimal functioning 5 Software Installation 5 1 The AbraScan RDS 1500 test strip reader uses Windows Mobile Device Center software or Microsoft Active Sync software to connect the reader to a desktop or laptop computer The software which will be utilized will depend on the version of Windows running on the computer and on whether internet access is available Note In order to init...

Страница 3: ...ired option 6 3 Enter can be used to run a test when at the Main Menu 6 4 The Left Side Button on the reader allows the user to go back to a previous screen 7 Keyboard The AbraScan RDS 1500 interface features an onscreen keyboard allowing the user to input data into the device by touching the desired letters and numbers using the stylus 7 1 Select the Keyboard Icon when it is present on the screen...

Страница 4: ...the unit to show the ratio of the intensity of the test line to the intensity of the control line 8 3 1 3 When checked Display European Date Format will display the date in the Day Month Year format 8 3 1 4 When checked Auto print will enable automatic printing of the results for each test that is run Note A printer must be connected to the unit to use this feature 8 3 2 To add edit or delete user...

Страница 5: ...st 9 2 4 Schemes allows the user to determine whether the unit will run on AC power or battery power and to change settings for time frames after which the unit switches to user idle system idle or suspends 9 3 The Help icon features tips and key information for any particular screen or function When navigating to any particular screen select the Help icon to get assistance in using the features o...

Страница 6: ...ests run on the unit reaches the specified threshold and calibration is necessary the reader will display a notification message in the test results prompting the user to calibrate the unit The need for calibration will also be indicated on the System Diagnostics screen If the device is due for calibration Calibration Status will show Out of Date and Next Calibration Due and Tests Remaining will s...

Страница 7: ... reader remove the previous test strip and replace with the new test strip to be read A sample ID for the new strip to be read can be entered by selecting New Sample ID entering the appropriate sample ID then pressing OK The test strip can then be read step 12 6 above Note If using only time and date information for sample identification no sample ID test strips can be read in immediate succession...

Страница 8: ...C using the USB cable provided 15 3 Activate the program by selecting the RDS 1500 Data Manager icon on the desktop 15 4 A Tool Bar is located at the top of the Data Manager screen Select Sync from the toolbar Results will then be copied from the reader results will not be deleted from the reader and will appear in both the reader and the PC until deleted from the reader 15 5 The previous 14 days ...

Страница 9: ...files list or the data currently displayed 15 9 4 Options Use to access additional options such as displaying the data format as dd mm yy instead of mm dd yy 15 9 5 Excel Use to export the current data file to a Microsoft Excel format 16 Abraxis Cyanotoxin Strip Test Kits Microcystins Strip Test for Finished Drinking Water PN 520016 5 Test and PN 520017 20 Test For use with treated drinking water ...

Страница 10: ......

Отзывы: