background image

Pressure  Simulator  is  ready  for  use.  The 

unit  has  been  factory  calibrated  for  use 

with  the  accessories  included  in  the  kit. 

No further calibration adjustments should 

be  neccessary  at  this  time.  If  the  unit  is 

to  be  used  with  a  sphygmomanometer 

other than that supplied with the unit, or 

if recalibration is necessary at a later date, 

then  see  the  section  titled  “Calibration 

Procedures”.

Familiarizing Yourself with the 

NASCO 

Life/form

®

 Blood Pressure 

Simulator Control Panel

 

Under  the  display  window  are  three 

buttons: Menu, Gap, and Calibrate. 

(See 

figure 5.)

 The systolic pressure is set by 

pressing  the  Menu  key  once.  The  pres-

sure is adjusted up or down using the up 

or  down  arrow  keys. 

(See  figure  6.)

 

The  diastolic  pressure  is  set  by  pressing 

the  Menu  key  a  second  time.  Adjust  the 

setting up and down with the arrow up or 

down  keys. 

(See  figure  7.)

  The  heart 

rate  is  set  by  pressing  the  Menu  key  a 

third time,

and adjusting the rate with the arrow up 

and  down  keys.  The  pulse  rate  can  be 

set from 0 beats per minute to 300 beats 

per minute. 

(See figure 8.)

 The palpa-

tion can be set to either on or pulseless. 

When  the  pulseless  setting  is  used,  the 

diastolic and systolic pressures will auto-

matically be set to 0.

Calibration Procedures

 

To  calibrate  the  simulator,  set  the 

unit up as described in the section titled 

“Using  The  NASCO 

Life/form

®

  Blood 

Pressure  Simulator.”  Apply  the  cuff  to 

the  simulated  arm.  Set  the  control  box 

systolic  pressure  to  150  mmHg  and 

set  the  diastolic  pressure  to  70  mmHg. 

Proceed with a simulated blood pressure 

measurement.  Note  the  discrepancy  in 

the  readings  between  the  gauge  and 

the  control  box.  Set  the  systolic  “offset.” 

Example:

  If  the  blood  pressure  read-

ing was taken and the sounds started at 

148 mmHg, then the offset is +2. If the 

sounds  started  at  152  mmHg,  the  offset 

is –2. For this example, assume that the 

sounds started at 148 mmHg. Press and 

hold the Calibration key until the systolic 

correction window appears. 

(See figure 

12.)

  Using  the  arrow  down  key  set  the 

correction to +2. Set the diastolic offset. 

Example:

 If the blood pressure reading 

was taken and the sounds stopped at 72 

mmHg, the offset is –2; or if the sounds 

stopped  at  68  mmHg,  the  offset  is  +2. 

For this example, assume that the sounds 

stopped  at  72  mmHg.  From  the  systolic 

window,  press  the  Menu  key  to  change 

the  diastolic  window. 

(See  figure  13.)

 

With the arrow up key, set the correction 

to –2. Press the Menu key again and the 

message  “CALIBRATION  COMPLETE” 

will appear. The main menu window will 

be displayed.

Preparing Your Equipment for 

Use with the NASCO 

Life/form

®

 

Blood Pressure Simulator

 

To  adapt  your  sphygmomanom-

eter  for  use  with  the  simulator,  it  is  first 

necessary to obtain the luer fitting and the 

T-fitting included with the simulator, which 

is  also  available  through  your  NASCO 

catalog  sales  office.  Using  a  wire  cut-

ting pliers or similar instrument, carefully 

cut  the  pressure  line  of  the  sphygmoma-

nometer  about  2"  from  the  gauge. 

(See 

figure 14.)

Figure 3

Figure 5

Figure 12

Figure13

Figure14

Figure 4

2

5

Figure 6

Figure 7

Содержание LF01095U

Страница 1: ...084U Sigmoidoscopic Examination LF01087U Central Venous Cannulation LF01095U Blood Pressure Arm LF01108U Intraosseous Infusion Simulator LF01142U Auscultation Trainer LF01162U Venatech IV Trainer LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer LF03602U Adult Airway Management on Manikin LF03603U Adult Airway Management Head Only LF03609U Child Airway Management Trainer LF03610U C...

Страница 2: ...it face up on the work sur face and turn it on by pressing the on off switch on the top right of the unit See figure 2 Observe the display and verify that a readable display is present indicat ing proper battery installation Note The control box has a battery saving feature which will turn the unit off after about 8 10 minutes if no keys are used within that period of time Take the T fitting and i...

Страница 3: ...ol box systolic pressure to 150 mmHg and set the diastolic pressure to 70 mmHg Proceed with a simulated blood pressure measurement Note the discrepancy in the readings between the gauge and the control box Set the systolic offset Example If the blood pressure read ing was taken and the sounds started at 148 mmHg then the offset is 2 If the sounds started at 152 mmHg the offset is 2 For this exampl...

Страница 4: ...e trainee or group of trainees an auxiliary blood pressure speaker amplifier system is available SB20146U If the auxiliary speaker is used the speaker is plugged into the EXT AUDIO AMP jack located next to the ARM output jack at the top of the unit Adjust the volume control to increase the output of the auxiliary speaker amplifier and proceed with the blood pressure measurement sequence as it woul...

Отзывы: