ColorSave 1000ST-INJ user manual
33
5.
M
AINTENANCE
5.1 Password
Some screens are password protected, such as calibration or
IP Address. You will see this screen while pressing them:
The password is 11
5.2 Static Calibration
Static calibration is the procedure that ensures that the weighing system is calibrated.
Static calibration should be performed under the following circumstances:
a)
Upon initial installation of the equipment.
b)
Any time the unit is dismantled and reinstalled.
c)
After replacing the load cell or the controller.
d)
Periodically, according to the company’s QA policy, but at least once a year.
The procedure for static calibration is described in Appendix B: Static Calibration.
Note: Check the calibration periodically by inserting a standard weight inside the hopper (no need to
empty it), and verifying that the change in the weight value matches the added weight. If there is a
deviation, a static calibration procedure should be performed again.
5.3 Dynamic Calibration
Dynamic calibration is used to update the gram per rotation value.
The procedure for dynamic calibration is described in Appendix C: Dynamic Calibration.