For an eight port multiplexer connect the pump
outlet with a single section of Bev-a-line tubing to
the lower right hand solenoid manifold.

9. There are two ways to configure the LI-8150 to vent

exhaust from the flasks.
a. Method 1: Dedicate a port on the multiplexer to

supply air to the auxiliary pump and expel
exhaust from the chambers.

i.

Disconnect the two urethane hoses

coming from one set of port connections on
the side of the LI-8150 from the solenoid
manifolds.
ii. Insert 1/4 inch quick connect plugs (part
#300-08151) into the open fittings on the
solenoid manifolds (Figure 8).
iii. Connect the urethane hose from the “IN”
connection of the port to the inlet of the
auxiliary pump using a quick connect
straight union.
iv. For a sixteen port model install a quick-
connect “Y” fitting (part #300-03367) on the
end of the “OUT” hose. For an eight port
multiplexer, install a quick connect straight
union (part #300-03123) on the “OUT” hose.

v. Cut two 35.5 cm lengths of Bev-a-line
tubing and connect the upper manifolds to
the “Y” fitting with each length of tubing.
For an eight port model use one length of
tubing to connect the upper right hand
manifold to the “OUT” hose.

b. Method 2: If dedicating a port to vent chamber

exhaust and supply air to the auxiliary pump is
undesirable, the supply/exhaust lines can be
routed out of the LI-8150 by leaving the case
open. If using this method, a filter (part #301-
08119) should be added before the pump inlet.

Constructing chambers

Flasks can be constructed from any gas tight containers,
provided they meet the volume considerations discussed
above and can be fit with connections for the LI-8150. Be
careful with plastic containers as sorption of water vapor
(and possibly CO

2

) by the container may affect the

measured fluxes. Here we describe how to use 1.89 L
glass canning jars (Ball Corporation, Daleville, IN) as
sample chambers.

1. With the lid removed from the jar, drill one 12 mm

hole off center in the lid.  Drill two 8 mm holes on

Figure 8.

 A sixteen port multiplexer configured for flask measurements. Port 16 has been re-purposed to vent flask exhaust

from the LI-8150 case and supply ambient air to the pump.

9981-108 power cable in an
unmodified multiplexer

To disconnect power to the pump,
unplug this connector

Exhaust lines

Install plugs (300-
08151) in the open
connectors on the
solenoid manifolds

Connect the exhaust
lines to the “out” line
from one of the ports
using part #300-03367

Connect the “in” line
from the same port
to the inlet on the 
auxiliary pump

Ambient air line 
from pump

Pump

6

Содержание LI-8100A

Страница 1: ...early in half because the increased volume serves to dilute the change in CO2 mole fraction within the system As dC dt decreases the potential for error in the measurement due to instrument noise incr...

Страница 2: ...s the mechanism by which CO2 and water vapor are exported from the system between measurements preventing condensation and excessively high CO2 mole fractions from developing in the flask The flushing...

Страница 3: ...o brass plugs from the right hand upper and lower solenoid manifolds 2 Using an adjustable wrench replace the three brass plugs with quick connect elbows part 300 07474 and the one steel plug with a q...

Страница 4: ...for flask measurements bottom right The flow path used during measurements is shown with blue arrows and the flow path used to flush flasks between measurements is shown with red arrows In both diagra...

Страница 5: ...ver plate from the instrument by loosening the two thumb screws Figure 7A Mount the pump in the LI 8150 case by attaching the mounting plate to one of the two threaded holes normally used to attach th...

Страница 6: ...thod 2 If dedicating a port to vent chamber exhaust and supply air to the auxiliary pump is undesirable the supply exhaust lines can be routed out of the LI 8150 by leaving the case open If using this...

Страница 7: ...ments are made at near ambient tempera tures a measurement of chamber air temperature may not be required Figure 9 Mason jar lid fitted with plumbing connections and a soil temperature thermistor Two...

Страница 8: ...as measured with a thermistor connected to V2 A volume correction can be made for the sample volume by entering a negative value for Extension Tube Volume cm3 Soil CO2 Flux System and LI 8150 Multiple...

Страница 9: ...on air trapped in the measurement loop from the previous measurement with the air in the flask Figure 12 The change in CO2 mole fraction with time calculated for various deadband lengths using the dat...

Страница 10: ...lated soil samples Jones and Kielland 2002 Haney et al 2008 and monitoring carbon fluxes from attached leaves and fruit Araki et al 1998 Kitano et al 1997 References Araki T M Kitano M Hamakoga and H...

Страница 11: ...8100A 8150 770 01 8150 661 8150 203 8150 670 16 or 8150 670 8 196 10534 210 01958 1 9981 173 9981 177 9881 204 140 04315 8150 250 300 07124 300 07125 300 07126 300 07127 167 07256 300 08118 300 07474...

Страница 12: ...licor com envsupport UK licor com LI COR Biosciences Global Headquarters 4647 Superior Street Lincoln Nebraska 68504 Phone 1 402 467 3576 Toll free 800 447 3576 Fax 1 402 467 2819 envsales licor com...

Отзывы: