Section 6.
Maintenance
The LI-7815 will require little maintenance. There are no user-serviceable parts in
the instrument except for the air inlet filter, air dry-down chemical, and pump. Do
not attempt to disassemble the instrument or conduct any maintenance procedures
other than those described here.
Instrument calibration
To adjust the instrument calibration, you can set the zero and span. The span is like
a slope (the
m
in
y
=
mx
+
b
). The zero is a spectral baseline. Adjusting the zero
changes every measurement (the relationship is non-linear). Adjusting the span
changes the slope. Calibration gases should be at a pressure between ambient and 35
kPa (5 PSI) above ambient pressure. We recommend installing a >30 cm vent before
the air inlet to vent the excess pressure and prevent diffusion of ambient air into the
gas stream.
Figure 6-1. Plumbing configuration for calibration gases. Use a low pressure regulator to
reduce pressure and a 30 cm vent to are recommended avoid pressurizing the inlet.
6-1
LI-7815 CO₂/H₂O Gas Analyzer
Содержание LI-7815
Страница 1: ...LI 7815 CO2 H2O Trace Gas Analyzer Instruction Manual...
Страница 2: ......
Страница 10: ...viii...
Страница 30: ...Section 2 Getting started 2 14 LI 7815 CO H O Gas Analyzer...
Страница 38: ...Section 3 Applications of the LI 7815 3 8 LI 7815 CO H O Gas Analyzer...
Страница 42: ...Section 4 Software overview 4 4 LI 7815 CO H O Gas Analyzer...
Страница 54: ...Section 4 Software overview 4 16 LI 7815 CO H O Gas Analyzer...
Страница 62: ...Section 5 Troubleshooting 5 8 LI 7815 CO H O Gas Analyzer...
Страница 70: ...Section 6 Maintenance 6 8 LI 7815 CO H O Gas Analyzer...
Страница 88: ...Appendix B Additional safety information B 6 LI 7815 CO H O Gas Analyzer...
Страница 92: ...Appendix C Specifications C 4 LI 7815 CO H O Gas Analyzer...
Страница 97: ......