LI-COR LI-6800 Скачать руководство пользователя страница 2

7

Slip the cable into the head cable bundle and press the
cable into the clasp.

8

Power the instrument back on.

9

Verify that the chamber is sealed and that the
software recognizes it.

Do the leak test under

Startup > System Tests > Chamber

Leak

. Check to see that the chamber is recognized

under

Constants > System Constants > ChType

.

Warning:

The light sources for this

product can emit potentially
hazardous optical radiation (RG-2
CAUTION POSSIBLY
HAZARDOUS OPTICAL RADIATION EMITTED FROM
THIS PRODUCT), in excess of the Exempt Risk Group.

Potential risk depends upon how one uses and installs
this product. Do not operate the light source while
detached from the chamber.

Do not look directly into

the light source under any circumstances.

Operate the

light sources with direct access to ambient air for cooling.
Do not use the product in any manner not described in
the manual.

LI-COR Biosciences

Global Headquarters

4647 Superior Street
Lincoln, Nebraska 68504
Phone: +1-402-467-3576 • Toll free: 800-447-3576 • Fax: +1-402-467-2819
[email protected] • [email protected] • www.licor.com/env

Regional Offices

LI-COR GmbH, Germany

Serving Andorra, Albania, Cyprus, Estonia, France, Germany, Iceland, Latvia, Lithuania,
Liechtenstein, Malta, Moldova, Monaco, San Marino, Ukraine, and Vatican City.

LI-COR Biosciences GmbH
Siemensstraße 25A • 61352 Bad Homburg
Germany
Phone: +49 (0) 6172 17 17 771 • Fax: +49 (0) 6172 17 17 799
[email protected] • [email protected]

LI-COR Ltd., United Kingdom

Serving Denmark, Finland, Ireland, Norway, Sweden, and UK.

LI-COR Biosciences UK Ltd.
St. John’s Innovation Centre
Cowley Road • Cambridge • CB4 0WS
United Kingdom
Phone: +44 (0) 1223 422102 • Fax: +44 (0) 1223 422105
[email protected] • [email protected]

LI-COR Distributor Network:

www.licor.com/env/distributors
© LI-COR, Inc. 05/16 984-16140

Отзывы: