Guided Tours
Tour #3: Controlling Chamber Conditions
3-36
Using the LI-6400
3
Temperature Control
Temperature control has two target options: a constant block temperature (the
block is the metal block that encompasses the sample and reference cells of
the IRGA), or constant leaf temperature.
has details, but the next two experiments illustrate the differences.
Experiment #7
Controlling on Block Temperature
This is the more direct and stable of the temperature control options.
1
Determine the ambient temperature.
Display line
h
of the default display map has block, air, and leaf temperatures.
Note the block temperature.
2
Set block target to 3¡ cooler than current value
Press
2
f4
B
<
value
- 3>
enter
(where
value
is the block temperature from Step
1). The external fans on either side of the chamber should start to run. Watch
the block temperature slowly drop to this new target value.
3
Note the temperature gradient
You should see that
Tblock
<
Tair
<
Tleaf
.
4
Now make the target 3 degrees warmer than ambient
Press
f4
<
value
+ 3>
enter
. The block temperature will rise a bit faster to this
new target. Heating is always more efficient than cooling.
5
Note the temperature gradient
You should see that
Tblock
>
Tair
>
Tleaf
.
Points to Remember
¥ Controlling on block temperature is slow but steady.
¥ Limit of control is generally within 7 degrees of ambient.
¥ The further
Tblock
is from ambient, the larger the temperature gradient
through the leaf chamber and IRGA.
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...